It is obvious that measurement of low object temperatures are more critical than
measuring high temperatures since the ‘disturbing’ radiation sources are relatively
much stronger in the first case. Should also the object emittance be low, the situation
would be still more difficult.
We have finally to answer a question about the importance of being allowed to use
the calibration curve above the highest calibration point, what we call extrapolation.
Imagine that we in a certain case measure
U
tot
= 4.5 volts. The highest calibration
point for the camera was in the order of 4.1 volts, a value unknown to the operator.
Thus, even if the object happened to be a blackbody, i.e.
U
obj
= U
tot
, we are actually
performing extrapolation of the calibration curve when converting 4.5 volts into tem-
perature.
Let us now assume that the object is not black, it has an emittance of 0.75, and the
transmittance is 0.92. We also assume that the two second terms of Equation 4 amount
to 0.5 volts together. Computation of
U
obj
by means of Equation 4 then results in
U
obj
= 4.5 / 0.75 / 0.92 – 0.5 = 6.0. This is a rather extreme extrapolation, particularly when
considering that the video amplifier might limit the output to 5 volts! Note, though,
that the application of the calibration curve is a theoretical procedure where no elec-
tronic or other limitations exist. We trust that if there had been no signal limitations in
the camera, and if it had been calibrated far beyond 5 volts, the resulting curve would
have been very much the same as our real curve extrapolated beyond 4.1 volts, pro-
vided the calibration algorithm is based on radiation physics, like the FLIR Systems
algorithm. Of course there must be a limit to such extrapolations.
21
198
Publ. No. 1557954 Rev. a155 – ENGLISH (EN) – February 7, 2006
21 – The measurement formula
Содержание ThermaCAM P65
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 7: ...ThermaCAM P65 User s manual Publ No 1557954 Rev a155 ENGLISH EN February 7 2006...
Страница 10: ...Reg No Status Designation 29 233 400 Pending U S x Publ No 1557954 Rev a155 ENGLISH EN February 7 2006...
Страница 16: ...xvi Publ No 1557954 Rev a155 ENGLISH EN February 7 2006...
Страница 28: ...INTENTIONALLY LEFT BLANK 4 12 Publ No 1557954 Rev a155 ENGLISH EN February 7 2006 4 Packing list...
Страница 32: ...INTENTIONALLY LEFT BLANK 5 16 Publ No 1557954 Rev a155 ENGLISH EN February 7 2006 5 System overview...
Страница 80: ...INTENTIONALLY LEFT BLANK 8 64 Publ No 1557954 Rev a155 ENGLISH EN February 7 2006 8 Tutorials...
Страница 98: ...INTENTIONALLY LEFT BLANK 9 82 Publ No 1557954 Rev a155 ENGLISH EN February 7 2006 9 Camera overview...
Страница 142: ...INTENTIONALLY LEFT BLANK 10 126 Publ No 1557954 Rev a155 ENGLISH EN February 7 2006 10 Camera program...
Страница 144: ...INTENTIONALLY LEFT BLANK 11 128 Publ No 1557954 Rev a155 ENGLISH EN February 7 2006 11 Folder and file structure...
Страница 150: ...INTENTIONALLY LEFT BLANK 12 134 Publ No 1557954 Rev a155 ENGLISH EN February 7 2006 12 Electrical power system...
Страница 154: ...INTENTIONALLY LEFT BLANK 14 138 Publ No 1557954 Rev a155 ENGLISH EN February 7 2006 14 Maintenance cleaning...
Страница 210: ...INTENTIONALLY LEFT BLANK 20 194 Publ No 1557954 Rev a155 ENGLISH EN February 7 2006 20 Theory of thermography...
Страница 245: ......