16.2.1
Finding the emissivity of a sample
16.2.1.1
Step 1: Determining reflected apparent temperature
Use one of the following two methods to determine reflected apparent temperature:
16.2.1.1.1
Method 1: Direct method
Action
Step
Look for possible reflection sources, considering that the incident angle = reflection
angle (a = b).
10588903;a1
Figure 16.1 1 = Reflection source
1
If the reflection source is a spot source, modify the source by obstructing it using
a piece if cardboard.
10589103;a2
Figure 16.2 1 = Reflection source
2
16
110
Publ. No. 1558407 Rev. a155 – ENGLISH (EN) – February 6, 2006
16 – Thermographic measurement techniques
Содержание ThermaCAM E320
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 7: ...ThermaCAM E320 User s manual Publ No 1558407 Rev a155 ENGLISH EN February 6 2006...
Страница 10: ...Reg No Status Designation 29 233 400 Pending U S x Publ No 1558407 Rev a155 ENGLISH EN February 6 2006...
Страница 16: ...INTENTIONALLY LEFT BLANK 1 2 Publ No 1558407 Rev a155 ENGLISH EN February 6 2006 1 Warnings cautions...
Страница 26: ...INTENTIONALLY LEFT BLANK 4 12 Publ No 1558407 Rev a155 ENGLISH EN February 6 2006 4 Packing list...
Страница 28: ...INTENTIONALLY LEFT BLANK 5 14 Publ No 1558407 Rev a155 ENGLISH EN February 6 2006 5 System overview...
Страница 70: ...INTENTIONALLY LEFT BLANK 8 56 Publ No 1558407 Rev a155 ENGLISH EN February 6 2006 8 Tutorials...
Страница 92: ...INTENTIONALLY LEFT BLANK 10 78 Publ No 1558407 Rev a155 ENGLISH EN February 6 2006 10 Camera program...
Страница 100: ...INTENTIONALLY LEFT BLANK 12 86 Publ No 1558407 Rev a155 ENGLISH EN February 6 2006 12 Maintenance cleaning...
Страница 142: ...INTENTIONALLY LEFT BLANK 18 128 Publ No 1558407 Rev a155 ENGLISH EN February 6 2006 18 Theory of thermography...
Страница 169: ......