Action
Step
Write down the emissivity.
12
➲
Please note the following:
■
Avoid forced convection
■
Look for a thermally stable surrounding that will not generate spot reflections
■
Use high quality tape that you know is not transparent, and has a high emissivity
you are certain of
■
This method assumes that the temperature of your tape and the sample surface
are the same. If they are not, your emissivity measurement will be wrong.
16.3
Reflected apparent temperature
This parameter is used to compensate for the radiation reflected in the object. If the
emissivity is low and the object temperature relatively far from that of the reflected it
will be important to set and compensate for the reflected apparent temperature cor-
rectly.
16
Publ. No. 1558439 Rev. a156 – ENGLISH (EN) – February 28, 2006
107
16 – Thermographic measurement techniques
Содержание TermaCAM E45
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 7: ...ThermaCAM EX300 User s manual Publ No 1558439 Rev a156 ENGLISH EN February 28 2006...
Страница 10: ...Reg No Status Designation 29 233 400 Pending U S x Publ No 1558439 Rev a156 ENGLISH EN February 28 2006...
Страница 16: ...INTENTIONALLY LEFT BLANK 1 2 Publ No 1558439 Rev a156 ENGLISH EN February 28 2006 1 Warnings cautions...
Страница 26: ...INTENTIONALLY LEFT BLANK 4 12 Publ No 1558439 Rev a156 ENGLISH EN February 28 2006 4 Packing list...
Страница 28: ...INTENTIONALLY LEFT BLANK 5 14 Publ No 1558439 Rev a156 ENGLISH EN February 28 2006 5 System overview...
Страница 86: ...INTENTIONALLY LEFT BLANK 10 72 Publ No 1558439 Rev a156 ENGLISH EN February 28 2006 10 Camera program...
Страница 94: ...INTENTIONALLY LEFT BLANK 12 80 Publ No 1558439 Rev a156 ENGLISH EN February 28 2006 12 Maintenance cleaning...
Страница 136: ...INTENTIONALLY LEFT BLANK 18 122 Publ No 1558439 Rev a156 ENGLISH EN February 28 2006 18 Theory of thermography...
Страница 162: ...148 Publ No 1558439 Rev a156 ENGLISH EN February 28 2006...
Страница 163: ......