![FLIR MR277 Скачать руководство пользователя страница 59](http://html1.mh-extra.com/html/flir/mr277/mr277_user-manual_2306037059.webp)
14.5
Safety Specifications
General Safety
CE/EN/UL/CSA/PSE 61010
Environmental Safety
REACH Regulation EC 1907/2006
RoHS2 Directive 2011/65/EC
WEEE Directive 2012/19/EC
JIS C 6802:2011 laser directive
IEC 60825-1 class II laser directive
FDA Laser directive
Encapsulation
IP54 (IEC 60529) with bottom flap fully sealed
Drop-proof
Designed to 6.6 ft. (2 m)
EMC
EN 61000-6-3
EN 61000-6-2
FCC 47 CFR Part 15 Class B
14.6
MR13 Sensor Measurement and Calculation Specifications
Parameter
Range
Accuracy
Resolution
Relative Humidity
10-90%
±2.5%
0.1
Air Temperature
32 to 122
℉
(0 to 50
℃
)
±1.1
℉
(±0.6
℃
)
Dew Point
-22 to 122
℉
(-30 to 50
℃
)
n/a (calculation)
Vapor Pressure
0.0–12.0 kPa
n/a (calculation)
Mixing Ratio
0-560 GPP
(0.0-80.0 g/kg)
n/a (calculation)
NOTE
Relative Humidity Sensor Note
:
The meter should be stored in an environment with a relative humidity similar to the area
to be tested. If the storage humidity differs by more than 50% RH from the area under test,
an acclimation period of up to 24 hours may be required to meet the specified RH
accuracy.
#NAS100005; r. AG/62027/62027; en-US
51
Содержание MR277
Страница 1: ...USER MANUAL Moisture Meter MSX IR Camera Hygrometer Model MR277...
Страница 2: ......
Страница 3: ...USER MANUAL Moisture Meter MSX IR Camera Hygrometer NAS100005 r AG 62027 62027 en US iii...
Страница 4: ......
Страница 8: ......
Страница 70: ...NAS100005 r AG 62027 62027 en US 62...
Страница 71: ......