18 MM560A-MM570A-EU-EN
V1.1 8/12
6) SPECIFICATIONS
Display:
4-4/5 digits 50,000 counts. Selectable stable mode
5-4/5digits 500,000 counts for DC Voltage, & 6 digits
999,999 counts for Hz
Polarity:
Automatic
Update Rate:
4-4/5 digits fast mode: 5 per second nominal;
5-4/5 digits stable mode: 1.25 per second nominal;
42 Segments Bar graph:
60 per second max
Operating Temperatur
e:
0
C to 45
C
Relative Humidity:
Maximum relative humidity 80% for temperature up to
31
C decreasing linearly to 50% relative humidity at 45
C
Pollution degree:
2
Storage Temperature:
-20
C to 60
C, < 80% R.H. (with battery removed)
Altitude:
Operating below 2000m
Temperature Coefficie
nt:
nominal 0.1 x (specified accuracy)/
C @(0
C -- 18
C or
28
C -- 40
C), or otherwise specified
Sensing:
AC, AC+DC True RMS
Safety:
Double insulation per IEC61010-1 2nd Ed., EN61010-1
2nd Ed., UL61010-1 2nd Ed. & CAN/CSA C22.2 No.
61010.1-0.92 to Category III 1000V AC & DC and
Category IV 600V AC & DC
MM560A Terminals
(to COM) Measurement Category:
V : Category III 1000 Vac & Vdc and Category IV 600 Vac
&
Vdc.
A / mA
A : Category III and Category IV 600 Vac and 300
Vdc.
MM570A Terminals
(to COM) Measurement Category:
V / A / mA
A : Category III 1000 Vac & Vdc and Category
IV 600 Vac & Vdc.
Overload Protection:
MM560A:
A & mA : 1A/600V, IR 10kA or better, F fuse (BBS-1)
A : 10A/600V, IR 100kA or better, F fuse (KTK-10)
V : 1050Vrms, 1450Vpeak
mV,
, & Others : 600 Vdc & Vac rms
MM570A:
A & mA : 0.44A/1000V, IR 10kA or better, F fus
e
(DMM-B-44/100)
A : 11A/1000V, IR 20kA or better, F fuse (DMM-B-11A)
V,
mV,
, & Others : 1050Vrms, 1450Vpeak
Transient protection
:
8kV (1.2/50
s surge)
E.M.C.:
Meets EN61326-1:2006 (EN55022, EN61000
-3-2,
EN61000-3-3, EN61000-4-2, EN61000-4-3,
EN61000-4-4, , EN61000-4-5, EN61000-4-6,
EN61000-4-8,
EN61000-4-11)
In an RF field of 3V/m
Capacitance function is not specified, other function
ranges: Total Accuracy = Specified Ac 100 digits,
performance above 3V/m is not specified
Содержание Extech Instruments MultiMaster MM560A
Страница 1: ...User s Guide MultiMaster Digital MultiMeter Series With PC Interface Models MM560A MM570A A...
Страница 12: ...12 MM560A MM570A EU EN V1 1 8 12 other readings indicate the diode is resistive or shorted defective...
Страница 24: ...24 MM560A MM570A EU EN V1 1 8 12 ISO 9001 Certified...