FUSE REPLACEMENT
1.
Disconnect the test leads from the meter.
2.
Remove the 2 Phillips head screws located on the back of the instrument and remove
the battery cover.
3.
Gently remove the fuse(s) and install the new fuse(s) into the holder(s).
4.
Always use fuses of the proper size and value (200mA/660V ceramic fast blow for the
mA / µA ranges, 10A/250V ceramic fast blow for the A range).
5.
Secure the fuse/battery compartment cover.
WARNING
: To avoid electric shock, do not operate your meter until the fuse cover is in
place and fastened securely.
Range Specifications
Function Range
Resolution Accuracy
200mV 0.1mV
2000mV 1mV
20V 0.01V
±
(0.5% r 2 digits)
200V 0.1V
DC Voltage
(V DC)
600V 1V
±
(0.8% r 2 digits)
200V 0.1V
AC Voltage
(V AC)
600V 1V
±
(1.2% r 10 digits
(50/60Hz)
200mA
100
μ
A
±
(1.2% r 2 digits)
DC Current
(A DC)
10A 10mA
±
(2.0% r 2 digits)
200
Ω
0.1
Ω
2000
Ω
1
Ω
20k
Ω
0.01k
Ω
200k
Ω
0.1k
Ω
±
(0.8% r 2 digits)
Resistance
20M
Ω
10k
Ω
±
(1.5% r 2 digits)
9V 10mV
Battery Test
1.5V 1mV
±
(1.0% r 2 digits)
-20°C to 750°C
1°C
Temperature
-4°F to 1400°F
1°F
±
(3% of r5°C/9°F)
Notes
:
Accuracy specifications consist of two elements:
•
(% reading) – This is the accuracy of the measurement circuit.
•
(+ digits) – This is the accuracy of the analog to digital converter.
Accuracy is stated at 18
o
C to 28
o
C (65
o
F to 83
o
F) and less than 75% RH.