FLIR DM91 USER MANUAL Document Identifier: DM91-en-US_AB
35
Lo-Z Voltage
(Auto AC & DC Detection)
Range
OL Reading
Resolution
Accuracy
600.0V DC & AC
660.0V
0.1V
±(2.0% + 3D)
1000V DC & AC
1100V
1V
Input Impedance: about 3kΩ
Frequency Response: 45 ~ 1kHz (Sine Wave)
Overload Protection: AC/DC 1000V
DC mV
Range
OL Reading
Resolution
Accuracy
600.0mV
660.0mV
0.1mV
±(0.5% + 2D)
Input Impedance: 10MΩ
Overload Protection: AC/DC 1000V
AC mV
Range
OL Reading
Resolution
Accuracy
600.0mV
660.0mV
0.1mV
±(1.0% + 3D)
Frequency Response: 45 ~ 1kHz (Sine Wave)
Input Impedance: 10MΩ
Overload Protection: AC/DC 1000V
DC Current
Range
OL Reading
Resolution
Accuracy
60.00mA
66.00mA
0.01mA
±(1.0% + 3D)
400.0mA
660.0mA
0.1mA
6.000A
6.600A
0.001A
±(1.0% + 3D)
10.00A
20.00A
0.01A
The accuracy of measurements > 10A is unspecified.
Maximum measurement time: > 5A for max. 3 minutes with at least 20-minute rest time.
> 10A for max.30 seconds with at least 10-minute rest time.
Overload Protection: AC/DC 11A for A terminal. AC/DC 660mA for mA terminal.