FT-112(D) Panel Weighing Indicator, Technical Manual, Rev.1.0.0, May 2019
Page
148
of
170
B6
Unit indication
0
First Unit (power on unit)
1
Second Unit
B5
Power On Zero
0
Not power on zeroed
1
Zeroed with power on zero
B4
Preset Tare
0
Preset tare is passive
1
Preset tare is active
B3
Indication
0
Gross mode
1
Net mode
B2
Motion Detection
0
Stable
1
Unstable
B1
Actual Weight or
Dynamic Result
0
Weight is actual weight
1
Weight is dynamic result
B0
Dynamic Operation
0
Dynamic is inactive
1
Dynamic weight is calculating
Table 15-1
– ALL Status table
Calibration Status (always 32 bit integer)
1
st
Dword (input) descriptions when read command is ‘Calibration Status’. Refer to 2
nd
Dword of PLC Output to
FT-112(D) Panel Input
Bit Number
1
st
Dword (input) Description
B31 … B11
Not in use
B10
0
No Error
1
The Calibration DIP switch is not ‘On’ position.
- Check the calibration DIP switch.
B9
0
No Error
1
Scale unstable
- Wait until scale become stable
- Check grounding wiring
B8
0
No Error
1
Calibration load value entry Error
- Test weight is too small. Increase the weight
B7
0
No Error
1
Calibration Error
- Calibration loading is not enough
- Check test weight loading
- Check load cell connections
B6
0
No Error
1
Instrument cannot be calibrating
- Load cell signal is very low or too high
B5
0
No Error
1
Instrument cannot be calibrating
- Check load cell cable
- Re-energize the instrument
B4
0
No Error
1
ADC Error
- Re-energize the instrument
- If shown again, change the board.
B3
0
No Error
1
Calibration Timeout
- Restart calibration
B2
0
None
1
Span calibration in process …
B1
0
None
1
Zero calibration in process …
B0
0
Not ready for calibration
1
Ready for calibration
Table 15-2 - Calibration status