FT-112(D) Panel Weighing Indicator, Technical Manual, Rev.1.0.0, May 2019
Page
110
of
170
Fast Continuous Data Output
Fast continuous “indicated weight” data output can be used only for the instruments, which can communicate
fast. The output rate is related with the baud rate. Use higher baud rate for faster data rate. CR and LF can be
enabled. You may reduce the data transfer speed by increasing the delay between data output packages.
The data format of the fast-continuous data output is;
[STX][STATUS][SIGN][WEIGHT VALUE][CR][LF]
Examples:
S+000123.4
(weight is stable and 123.4)
D+000123.4
(weight is unstable and 123.4)
M+000123.4
(Dynamic weighing result is 123.4)
-------------
(Dynamic weighing in calculating)
+
(Over load)
-
(Under load)
O
(ADC out error)
Print Mode
The format of the data output in Print mode can be selected in 5 different type forms in the sub-block 16- .
Only
one
serial port can be programmed for printing. Print mode data outputs is sent by pressing the
key
or by receiving ASCII P command via serial port, or via Ethernet port by setting remote IP and remote port
number, refer to parameter 19E and 19F.
SINGLE LINE
You can transmit the printout data in single line format by pressing
key. This format is recommended to
send the print data to PC or to any host. The data output structure is;
12.02.2016 14:47 CN: 71 G: 3.007kg T: 1.001kg N: 2.006kg *ALB: 5*
MULTI LINE
You can send the data in multiple line formats as shown in the label given below by pressing
key. The
data output including can be programmed with printer parameters. The multiline data output can be
programmed for 16-byte narrow printers and for others.
Format for 16-character printer:
16F1
– Includes only Gross or
Net weight value
16F2
– Includes Gross or Gross\
Tare\Net weight values
DATE
TIME
ID1
Name
ID1
Data
ID2
Name
ID2
Data
CN
GROSS
PT or
TARE
NET
ALB
C
R
L
F
10
3
5
3
Max. 16 3
Max. 32 3
Max. 16 3
Max. 32 3
9
3
13
3
13
3
13
4
12
1
1
Flintec
www.flintec.com
Meckesheim, Germany
23.02.2019
09:16
CN 34
OPERATOR
JOSEPH SMITH
MATERIAL
ASCORBIC ASID
G 4.772 kg
T 1.675 kg
N 3.097 kg
* Thank you *
Flintec
www.flintec.com
Meckesheim, Germany
23.02.2019
09:16
CN 34
OPERATOR
JOSEPH SMITH
MATERIAL
ASCORBIC ASID
G 4.772 kg
* Thank you *
Header
Date & Time
Ticket no
Identifications
Gross/Tare/
Net weights
Footer