1. Before you can use the remote display FRD100, it must be
configured with the software “V100DGA201.exe”
First copy the file „V100DGA201.exe“ to your computer (e.g. own files) and double click on the file.
Start the programming after the connections described in topic 2 were done correctly.
Connect the connector X2 (10pins) with the computer:
Start the software
“V100DGA” on your PC / Laptop.
Chose the ComPort
The letter „P“ will appear if the communication is
correct.
2. Prepare the connections
And connect FRD100 to your PC
3. The remote display supports Flintec protocol for FAD-30, FAD-
40,0 FT-1x, FT-11x devices.
Flintec Protokoll:
<STX>Daten<CR><CHK>
Functions:
Decimal, Gross/Net, motion, positive/negative and overload
Status-LEDs
Serial port settings:
The settings at FRD and FT-1x must be the same:
Baudrat
– 9600
Parität
– 8 Bit, none
Settings in the software: (Flintec protocol)
Protocol
Custom 6
Special function
Function 1
connector
Protocoll
-
Number 026