FMS-2530 Tabletop Ultrasonic Scaler V.1
15
supply
input lines IEC
61000-4-11
70% UT(30% dip in
UT) for 25 cycles
<5% UT(>95% dip in
UT) for 5 s
70% UT(30% dip in
UT) for 25 cycles
<5% UT(>95% dip in
UT) for 5 s
continued operation during power
mains interruptions, it is
recommended that the FMS-2530
be powered from an
uninterruptible power supply.
Power
frequency(50/60 Hz)
magnetic field IEC
61000-4-8
3 A/m
3 A/m
The FMS-2530
contains no
devices susceptible to magnetic
fields.
NOTE: UT is the AC mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration-electromagnetic immunity
The FMS-2530 is intended for use in the electromagnetic environment specified below.
The customer or the user of the FMS-2530 should assure that is used in such and environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic environment-guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 V/m
150 KHz to 80
MHz
3 V/m
80MHz to 2,5
GHz
3 V/m
3 V/m
Portable and mobile RF communications
equipment should be used no closer to any part
of the FMS-2530
including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter. Recommended separation
distance:
d = 1,2 ˇP
d = 1,2ˇP 80MHz to 800 MHz
d = 2,3 ˇP 800MHz to 2,5 GHz
Where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in metres
(m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
(a) should be less than the compliance level in
each frequency range. (b) Interference may
occur in the vicinity of equipment marked with
the following symbol:
NOTE1