FMS-2530 Tabletop Ultrasonic Scaler V.1 

15 

 

supply 
input lines IEC 
61000-4-11 

70% UT(30% dip in 
UT) for 25 cycles 
<5% UT(>95% dip in 
UT) for 5 s 

70% UT(30% dip in 
UT) for 25 cycles 
<5% UT(>95% dip in 
UT) for 5 s 

continued operation during power 
mains interruptions, it is 
recommended that the FMS-2530 
be powered from an 
uninterruptible power supply. 

Power 
frequency(50/60 Hz) 
magnetic field IEC 
61000-4-8 

3 A/m 

3 A/m 

The FMS-2530

 

contains no 

devices susceptible to magnetic 
fields. 

NOTE: UT is the AC mains voltage prior to application of the test level. 
 
 
 
 
 

Guidance and manufacturer’s declaration-electromagnetic immunity 

The FMS-2530 is intended for use in the electromagnetic environment specified below. 
The customer or the user of the FMS-2530 should assure that is used in such and environment. 

Immunity test 

IEC 60601 test 

level 

Compliance 

level 

Electromagnetic environment-guidance 

Conducted RF 
IEC 61000-4-6 
Radiated RF 
IEC 61000-4-3 

3 V/m 
150 KHz to 80 
MHz 
3 V/m 
80MHz to 2,5 
GHz 

3 V/m 
 
3 V/m 

Portable and mobile RF communications 
equipment should be used no closer to any part 
of the FMS-2530

 

including cables, than the 

recommended separation distance calculated 
from the equation applicable to the frequency of 
the transmitter. Recommended separation 
distance: 
d = 1,2 ˇP 
d = 1,2ˇP 80MHz to 800 MHz 
d = 2,3 ˇP 800MHz to 2,5 GHz 
Where P is the maximum output power rating of 
the transmitter in watts (W) according to the 
transmitter manufacturer and d is the 
recommended separation distance in metres 
(m). 
Field strengths from fixed RF transmitters, as 
determined by an electromagnetic site survey, 
(a) should be less than the compliance level in 
each frequency range. (b) Interference may 
occur in the vicinity of equipment marked with 
the following symbol: 
NOTE1 

Содержание FMS-2530

Страница 1: ...ULTRASONIC SCALER FMS 2530 Flight Dental Systems 21 Kenview Blvd Unit 9 Ontario Canada L6T5G7 www flightdentalsystems com Toll Free 1 866 799 0517...

Страница 2: ...s 8 Installation Instructions 7 8 1 General Information 8 2 Water Line Requirements 8 3 Electrical Requirements 8 4 Unpacking the System 8 5 System Installation 8 6 Power Connection 8 7 Water Supply L...

Страница 3: ...leaving the office The use of an in line water filter is recommended Use the device for the intended purpose only The ultrasonic scaler must be used by qualified and trained personnel in a medical de...

Страница 4: ...able parts inside this device Please contact your dealer to inquire about repair service and technical support CAUTION EQUIPMENT IS NOT SUITABLE FOR USE IN THE PRESENCE OF FLAMMABLE ANESTHETIC MIXTURE...

Страница 5: ...ATURE 10 C 40 C RELATIVE HUMIDITY 30 75 ATOMOSPHERIC PRESSURE 700 kPa 1060kPa 5 4 Transportation And Storage Environment AMBIENT TEMPERATURE 10 C 70 C RELATIVE HUMIDITY 10 90 ATOMOSPHERIC PRESSURE 500...

Страница 6: ...imum operator and patient safety carefully follow section 12 system maintenance and care information detailed in the operating instructions As with high speed handpieces and other dental devices the c...

Страница 7: ...FMS 2530 Tabletop Ultrasonic Scaler V 1 7 infection control procedures...

Страница 8: ...ystem the dental office water line should be thoroughly flushed prior to connection to the System After flushing the system verify that there are no leaks 8 3 Electrical Requirements Refer to Section...

Страница 9: ...System to the insert AC Power Indicator Light Illuminates when the system is switched on Sterilizable Sleeve of Handpiece Holder Securely holds the Hand piece when system is not being used Water flow...

Страница 10: ...t Pedal Controls and Operation The Foot Pedal is a two position momentary switch which activates both Ultrasonic energy and water flow at the insert tip Foot Pedal Released Both ultrasonic activation...

Страница 11: ...f the handpiece over a sink or drain Check spray temperature to verify fluid is reaching the working end of the insert tip Adjust the water irrigatation control knob to ensure adequate flow for the se...

Страница 12: ...nfectant on a clean towel and wipe all surfaces Allow disinfectant solution to air dry Do not spray disinfectant solution directly on the ultrasonic scaler 3 Place a sterilized handpiece sleeve on the...

Страница 13: ...acle on the back of the unit 2 Check that the system s plug is fully seated in an appropriate AC receptacle and that AC current is present Power Indicator Light is lit 1 Check that the foot pedal conn...

Страница 14: ...pairs during the warranty period are not carried out by an authorized engineer dealer the warranty will expire immediately 16 3 Warranty Exception The warranty stated herein is the sole warranty appli...

Страница 15: ...electromagnetic immunity The FMS 2530 is intended for use in the electromagnetic environment specified below The customer or the user of the FMS 2530 should assure that it is used in such an environme...

Страница 16: ...environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Conducted RF IEC 61000 4 6 Radiated RF IEC 61000 4 3 3 V m 150 KHz to 80 MHz 3 V m 80MHz to 2 5 GHz...

Страница 17: ...The FMS 2530 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the FMS 2530 can help prevent electromagnetic interfere...

Страница 18: ...FMS 2530 Tabletop Ultrasonic Scaler V 1 17...

Отзывы: