Clave 23+ Instruction Manual v1.4 

Clave 23 + Quick Guide 

Set the Time and Date 

1.

 

Power on, wait, and then press Select. Press the Down button and highlight ‘Basic Set’. 
Enter ‘Basic Set’ by pressing Select. 

2.

 

Select item to change by pressing Select. 

3.

 

Adjust numbers by pressing Up and Down buttons. The time is a 24 hr. Clock. 

4.

 

Press Select at the last number to save and exit that menu. The Counter cannot be 
changed in this screen. 

5.

 

Power off. 

Retrieve Record on USB Stick 

1.

 

Power on, wait, and then press Select. Press Down button and highlight ‘Report’. Enter 
‘Report’ by pressing Select. A list of the last 20 cycles will be accessible. Insert a USB 
stick correctly into the USB connector. 

2.

 

Use the Up and Down keys to highlight the record you want. Press Select. If a printer is 
connected, it will also print a copy of that report. Wait at least 10 seconds for the report to 
be transferred to the USB stick. Repeat the procedure (highlight and select) for any other 
records to be transferred to the USB stick. 

Checking Maintenance Screen 

1.

 

Hold the Up and Down buttons while powering on. Let go when you enter the interface. 
Scroll to ‘Maintenance’ and highlight it. Press Select button. On a new install, use the 
select button to change the ‘End’ date to 1 yr in the future. Change the cycle count to the 
expected cycle count for a year or 1000. Use the up/down buttons to change the values. 
When finished, press the select button and power off. E88 or ‘wrench’ icon will come up 
when either value is reached. Don’t make changes on a unit already in use for a while. 

Changing Drying Time 

1.

 

Power on while pressing and holding the Select button for 5 seconds. This will enter the 
advanced settings mode. 

2.

 

Press the Select button to select Parameter. Press Select button to enter the menu. 
Highlight the program you want to change. Press Select button to move to the drying 
time. Use the up/down buttons to change the drying time. Press the Select button to save 
and return to the earlier menu. It is not recommended to change the default holding time. 

 

 

 

Содержание Clave 23+

Страница 1: ...1 Clave 23 Instruction Manual v1 4 Clave 23 Instruction Manual Steam Sterilizer Distributed by...

Страница 2: ...ode E88 will be displayed after 1 year of use or after a certain number of cycles whichever comes first This is programmed into the maintenance screen The Clave 23 is the same as a Clave 23 except it...

Страница 3: ...ials 8 5 3 Selecting the sterilization program 8 5 4 Running the sterilization program 9 5 5 Basic set 10 5 6 Record of the cycle 10 6 Advanced settings 11 7 Maintenance 14 8 Error codes 16 9 Transpor...

Страница 4: ...cedure highlight and select for any other records to be transferred to the USB stick Checking Maintenance Screen 1 Hold the Up and Down buttons while powering on Let go when you enter the interface Sc...

Страница 5: ...denser exhaust 6 Used water tank drain port 14 Rating plate 7 Door handle 15 Circuit breaker 8 Door 16 Power socket Security Notice For safe operation please pay close attention to the warning symbols...

Страница 6: ...imum level Operation temperature 5 C 40 C Exterior dimensions 480mm W x 445mm H x 700mm D 18 9 W x 17 5 H x 27 6 D Net weight 50kg 110 2 lbs Noise level 70db Relative humidity Maximum 80 non condensin...

Страница 7: ...1 Setup 5 1 1 Open the door and remove all of the inner contents 5 1 2 Connect the power cord to an outlet of the appropriate voltage 5 1 3 Power on The switch is located on the right side of the cove...

Страница 8: ...on to avoid water pooling Don t stack the trays one above the other or put them in direct contact with the walls of the sterilization chamber Always use the instrument tray handle Wrap the instruments...

Страница 9: ...chamber using the tray handle 5 4 1 After the instruments are loaded you may close and lock the door by turning the door handle clockwise until it stops Caution You must turn the door handle to the m...

Страница 10: ...ing time to N It will skip the drying stage 5 5 Basic Set 5 5 1 Highlight the Basic Set and press the SELECT button 5 5 2 Select the item by pressing SELECT button 5 5 3 Adjust the value by pressing t...

Страница 11: ...Ambient Pressure 6 2 2 Adjust the parameter by pressing button Press the SELECT button to save Press START button to cancel and exit 6 2 3 The value is 0 2 If the machine can t enter the holding time...

Страница 12: ...ht side panel The information will be automatically recorded to the USB drive after the cycle has completed The name of the file is determined by the serial number of the machine and the cycle number...

Страница 13: ...50 5kPa T4 13 31 31 084 0C 009 7kPa T4 15 11 50 083 5C 009 9kPa T5 13 35 42 109 8C 102 7kPa T5 15 15 50 108 2C 101 1kPa T6 13 36 22 098 8C 009 9kPa T6 15 16 30 097 2C 009 7kPa TS 134 8C 222 5kPa TS 00...

Страница 14: ...Replace the bacteria filter Every 6 12 months Replace the door seal 7 1 Clean the distilled water tank every week with isopropyl alcohol or a medical disinfectant Drain unused water 7 2 Clean the cha...

Страница 15: ...ighten the sealing plate 7 4 3 Turn the nut until the sealing plate is tight If the door knob is too tight you may also turn the nut clockwise to loosen it Caution Never adjust the chamber door while...

Страница 16: ...curs at 8 minutes E06 Door lock fault Make sure you have turned the door handle to the maximum position Check the door switch E09 Pressure leak takes too long to reach pressure and temperature Check i...

Страница 17: ...Action Interruption of the power to the transformer Contact a service person 3 Safety valve Protection against possible sterilization over pressure Action Releases steam and restores pressure inside...

Страница 18: ...mg L 1 0mg kg Silicon dioxide SiO2 1mg L 0 1mg kg Iron 0 2mg L 0 1mg kg Cadmium 0 005mg L 0 05mg kg Lead 0 05mg L 0 1mg kg Rest of heavy metals 0 1mg L 0 1mg kg Chloride 2mg L 0 1mg kg Phosphates 0 5m...

Страница 19: ...apped porous material 1 25 0 40 Single wrapped porous material 1 10 0 30 Dual wrapped porous material 0 75 0 25 Single wrapped hollow material 4 00 1 25 Dual wrapped solid and hollow material 2 00 0 6...

Страница 20: ...20 Clave 23 Instruction Manual v1 4...

Страница 21: ...ion Manual v1 4 APPENDIX 3 WIRING DIAGRAM TP1 Steam generator temperature sensor TP2 Inner temperature sensor TP3 Chamber wall temperature sensor V1 Air release valve normally open V4 Water release va...

Страница 22: ...22 Clave 23 Instruction Manual v1 4 Appendix 4 Hydraulic Diagram...

Страница 23: ...must be paid for by the owner of the Autoclave Flight will not accept COD shipments If repairs are needed during the first 90 days after purchase of an autoclave and a local authorized service dealer...

Страница 24: ...s regarding the document is welcomed For any comments and concerns please mail e mail or phone us at Sales and Marketing sales flightdentalsystems com 1 866 799 0517 905 799 0517 Technical Support 1 8...

Отзывы: