15 Outputs
82
UMFLUXUS_F5WDV1-1EN, 2016-04-08
15.6.2
Reset and Initialization of the Alarms
After an initialization, all alarm outputs will be initialized as follows:
Press key C three times during the measurement to set all alarm outputs to the idle state. Alarm outputs whose switching
condition is still met will be activated again after 1 s. This function is used to reset alarm outputs of the type
HOLD
if the
switching condition is no longer met.
By pressing key BRK, the measurement will be stopped and the main menu selected. All alarm outputs will be de-ener-
gized, independently of the programmed idle state.
15.6.3
Alarm Outputs During Transducer Positioning
At the beginning of the transducer positioning (bar graph display), all alarm outputs switch back to the programmed idle
state.
If the bar graph is selected during the measurement, all alarm outputs will switch back to the programmed idle state.
An alarm output of the type
HOLD
that has been activated during the previous measurement will remain in the idle state af-
ter the transducer positioning if the switching condition is no longer met.
Switching of the alarm outputs into the idle state will not be displayed.
15.6.4
Alarm Outputs During the Measurement
An alarm output with switching condition
MAX
or
MIN
will be updated max. once per second to avoid humming (i.e. fluctua-
tion of the measured values around the value of the switching condition).
An alarm output of the type
NON-HOLD
will be activated if the switching condition is met. It will be deactivated if the switch-
ing condition is no longer met. The alarm will remain activated min. 1 s even if the switching condition is met for a shorter
period of time.
Alarm outputs with the switching condition
QUANT.
will be activated if the limit is reached.
Alarm outputs with the switching condition
ERROR
will only be activated after several unsuccessful measuring attempts.
Therefore, typical short-term disturbances of the measurement (e.g., switching on of a pump) will not activate the alarm.
Alarm outputs with the switching con
→
- -
→
+ and the type
NON-HOLD
will be activated with each change of the flow
direction for approx. 1 s (see Fig. 15.2).
Alarm outputs with the switching condition
+
→
-
-
→
+
and the type
HOLD
will be active after the first change of the flow direc-
tion. They can be switched back by pressing key C three times (see Fig. 15.2).
If there is an internal adaptationto changing measuring conditions, e.g., to a considerable rise of the fluid temperature, the
alarm will not switch. Alarm outputs with the switching condition
OFF
will be set automatically to the switching function
NO
Cont.
.
Tab. 15.5: Alarm state after an initialization
func
OFF
typ
NON-HOLD
mode
NO Cont.
Limit
0.00
Fig. 15.2: Behavior of a relay when the flow direction changes
0
0
+
-
+
-
flow
type
NON-HOLD
type
HOLD
reset of the alarm
(3x key C)
approx. 1 s
flow
Содержание FLUXUS WD100
Страница 1: ...User Manual UMFLUXUS_F5WDV1 1EN Ultrasonic Flowmeter for Water FLUXUS WD100 FLUXUS WD200...
Страница 4: ...FLUXUS WD100 WD200 UMFLUXUS_F5WDV1 1EN 2016 04 08...
Страница 8: ...FLUXUS WD100 WD200 Table of Contents 8 UMFLUXUS_F5WDV1 1EN 2016 04 08...
Страница 103: ...D Declarations of Conformity FLUXUS WD100 WD200 UMFLUXUS_F5WDV1 1EN 2016 04 08 103 D Declarations of Conformity...
Страница 104: ...FLUXUS WD100 WD200 D Declarations of Conformity 104 UMFLUXUS_F5WDV1 1EN 2016 04 08...
Страница 105: ......
Страница 106: ......