97
10 Basic Measurement
10.7.4 Consistency Check
If a wide range for the sound speed has been entered in the program branch
PARAMETER
or the exact parameters of the medium are not known, a consistency check is recom-
mended.
The transducer distance can be displayed during measurement by scrolling with key
.
The optimum transducer distance is calculated on the basis of the measured sound
speed. It is therefore a better approximation than the first value which had been calculat-
ed on the basis of the approximate sound speed range entered in the program branch
PARAMETER
.
If the difference of optimum and entered transducer distance is less than specified in Ta-
ble 10.1 or Table 10.2, the measurement is consistent and the measured values are val-
id. The measurement can be continued.
If the difference is greater, adjust the transducer distance to the displayed optimum val-
ue. Afterwards, check the signal quality and the signal amplitude bar graph (see section
10.7.3). Press ENTER.
The optimum transducer distance (here: 50.0 mm) will be
displayed in parentheses in the upper line, followed by the
entered transducer distance (here: 54.0 mm). The latter
value must correspond to the actually adjusted transducer
distance. Press ENTER to optimize the transducer dis-
tance.
Table 10.1: Standard values for signal optimization of shear wave transducers
transducer frequency
difference between optimum and entered transducer distance [mm]
P
8
M
10
K
15
G
20
Table 10.2: Standard values for signal optimization of Lamb wave transducers
transducer frequency
difference between optimum and entered transducer distance [mm]
Q
-3...+5
P
- 6...+10
M
-10...+20
K
-25...+40
H
-35...+60
G
-50...+100
DISP
L=(50.0) 54.0 mm
54.5
m3/h
Содержание FLUXUS G704
Страница 4: ......
Страница 174: ...174 UMG70XV3 4EN 12 01 2009 FLUXUS G709 dimensions in mm 198 1 208 25 129 222 2 15 122 5 110 M2 5 43 7 213 129 170 5...
Страница 235: ...UMG70XV3 4EN 12 01 2009 235 D Certificates D Certificates...
Страница 236: ...236 UMG70XV3 4EN 12 01 2009 D Certificates...
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ...UMG70XV3 4EN 12 01 2009 D Certificates...
Страница 241: ......
Страница 242: ......
Страница 243: ......
Страница 244: ......
Страница 245: ......
Страница 246: ......
Страница 247: ......
Страница 248: ......
Страница 249: ......
Страница 250: ......
Страница 251: ......
Страница 252: ......
Страница 253: ......
Страница 254: ......
Страница 255: ......
Страница 256: ......
Страница 257: ......