background image

Page 

4

 of 

5

 

 

 

 

Entering the MEASURING Mode 

 

18)

 

Using the 

or 

keys, place the carrots around 

>MEA<

 

(

MEASURING

) then press 

Enter

 

19)

 

Check Mark 

(

) under a Channel indicates this channel 

will be making measurements. Use the 

and 

6

 keys to 

move the carrots to the Channel you plan on activating 
then use the 

and 

keys to change the minus sign (

-

) to a 

check mark (

 ). A dot (

.

) under any Channel indicates this 

Channel has not been programmed. 

 
20)

 

If you are recording data, this screen allows you to apply 
file numbers to the pipe you are measuring. For example, 
you can measure multiple pipes, assign different numbers 
to each pipe, and the data will be separated in memory by 
the 

Measuring Point Number

 
21)

 

Use the keypad to enter the 

Sound Path 

to indicate the 

placement of the transducers onto the pipe then press 

ENTER

Sound Path 1

 will place the transducers on 

opposite sides of the pipe. 

Sound Path 2

 places the 

transducers on the same side of the pipe. 

Sound Path 2 

is 

recommended and simpler to install. 

      

Sound Path 1 

has stronger communications signals. 

 
 
 
 
 
 
22)

 

Transd. Distance

 indicates the separation distance 

between the two transducers as placed on the pipe. 

You 

MUST mount the transducers with this separation

. Note 

that Negative Transducer Distance is possible in Sound 
Path 1 on small diameter pipes. 

 

 

 

 

CHANN:   >A<   B    Y    Z 
MEASUR    

     

   .    .

 

    . 

  

  par         >MEA<         opt           sf 

 

MEASURING 

 

A: Meas. Point No.: 
           xxx  (



 

    . 

A: Sound Path: 
                      2 

    . 

 

Transd. Distance 
 A:      1.12 inch Reflex 

    . 

 