9 Measurement
FLUXUS F401
9.5 Execution of special functions
75
UMFLUXUS_F401V2-1EN, 2020-06-12
9.5
Execution of special functions
Commands that can be executed during a measurement are displayed in the upper line.
A command begins with →. If programmed, a program code has to be entered first.
• Press key
until the command is displayed.
• Press ENTER.
The commands available are the following:
9.6
Determination of the flow direction
The flow direction in the pipe can be detected with the help of the displayed volumetric
flow rate sign in conjunction with the arrow on the transducers:
• The fluid flows in the direction of the arrow if the displayed volumetric flow rate is posi
-
tive (e.g., 54.5 m³/h).
• The fluid flows against the arrow direction if the displayed volumetric flow rate is nega
-
tive (e.g., -54.5 m³/h).
9.7
Stop of the measurement
The measurement is interrupted by pressing key BRK if it is not protected by a program
code, see section 18.4.
Tab. 9.6:
Executable commands during the measurement
command
explanation
→Adjust transd.
toggling to transducer positioning
If a program code is active, the measurement will be continued 8 s
after the last keyboard entry.
→Clear totalizer
reset of flow totalizers to zero
→Break measure
stop the measurement and return to the main menu
Notice!
Be careful not to stop a current measurement by inadvertently pressing key BRK.