INSTRUCTIONS 

 

1.

 

Insert the BioFlex

®

 Cell Seeder into a 

BioFlex

®

 Baseplate, similar to how the 25 

mm equibiaxial Loading Stations are placed 
in the baseplate. 

Optional:

 Add silicone 

lubricant to the upper rim of a Cell Seeder 
post to aid membrane gliding into the Cell 
Seeder body. 

2.

 

Place a BioFlex

® 

plate into its respective 

gasket positioned over the Cell Seeder, 
ensuring that the wells in the plate align 
concentrically over the posts on the Cell 
Seeder. 

3.

 

Using the FX-5000

 Tension System, create 

and start a regimen with the following 
settings: 

 

SHAPE: Static 

 

MIN: 0.0 

 

MAX: 15.0 

 

FREQ: 1 Hz (

this value must still be 

entered even though the regimen is 
static

 

DURATION: Equal to the seeding 
time (see step 5 below) plus the time 
needed to plate the cells into the 
well. 

 

PLATFORM: BFlx Loading Station 
(25mm)  

4.

 

The recommended suspension volume when 
using a 6-well, BioFlex

® 

Cell Seeder for an 

individual well is 1 mL medium plus cells. 

This volume is large enough to allow 
uniform distribution of the media-cell 
suspension over the growth surface area 
when dispensing cells and medium with a 
1000 µL micropipette. Also, this volume is 
small enough to retain the media-cell 
suspension within the boundary of the Cell 
Seeder form when handling the baseplate. 

5.

 

Due to the limited suspension volume, it is 
advised that the user observe the seeding 
time required for the cells to adhere to the 
membrane. A seeding time of at least two 
hours is recommended, but the required time 
is dependent on the type of cells being used.  

6.

 

After the cells have adhered to the substrate, 
remove the vacuum gradually. It is 
recommended to create a slow vacuum 
release regimen that reduces the vacuum a 
set percent every “n” seconds. An example 
regimen is outlined below in Table 1. 

7.

 

Slowly add 1 mL of fresh medium to each 
well to increase the volume for proper 
aspiration of non-adherent cells. Then, insert 
an aspirator tip near the side of each well 
(adjacent to the seeded cells) and aspirate 
the media containing non-adherent cells.   

8.

 

Add 3 mL of fresh media to each well and 
return the BioFlex

®

 plate to the incubator 

until you are ready to apply strain. 

 

Table 1.

 Sample regimen parameters for releasing the vacuum pressure slowly. 

Step Shape Min  Max  Freq  DC%  dd:hh:mm:ss  Back 

To  Repeat 

1

 

Static 

0.0 

15.0 

1.0

50.0

00:02:30:00

0

2

 Static 

0.0  13.0 1.0

50.0

00:00:00:06

0

3

 

Static 

0.0 

11.0 

1.0

50.0

00:00:00:06

0

4

 Static 

0.0  9.0  1.0

50.0

00:00:00:06

0

5

 

Static 

0.0 

7.0 

1.0

50.0

00:00:00:06

0

6

 

Static 

0.0 

5.0 

1.0

50.0

00:00:00:06

0

7

 

Static 

0.0 

3.0 

1.0

50.0

00:00:00:06

0

8

 

Static 

0.0 

1.0 

1.0

50.0

00:00:00:06

0

In this example, we recommend a two hour adherence time plus a 30 minute set-up time (Step 1). Following 
seeding, this example releases the strain by 1-2% every six seconds. 