Calibration procedure

In order to limit parameter handling, FKG recommends reprogramming the factory settings of the memory «

M6

» 

. To do 

this, please refer to the programming protocol on the front page of this document.

Before using directly after purchase, whenever the motor handpiece or contra angle has been replaced, or if the motor alternates 

between forward and reverse rotation outside the canal, it is recommended to calibrate the instrument in order to control and 

adjust the measurements. To do this please follow the instructions below:

1.  Put a commonly used file into the contra angle.

2.  Turn the unit on and make sure the battery is fully charged (three bars are displayed 

 ).

3.  Select memory M6 using the Plus or Minus switch 

.

4.  Set the rotation mode for “Rev”:

a.  Press the Select switch   and select the primary function “Rotation mode”. By pressing several times, you will display the 

primary functions.

b.  Press the Plus or Minus switch 

 to set the “Rev” rotation mode.

5.  Turn the unit off. To do this, simultaneously press the Select switch   and the Main switch  .

6.  Hold down the Plus and Minus switches 

 and turn the unit back on.

7.  When “CO-Adjst” appears in the display, press the Selct switch  . 

The motor will start running. Make sure there is no load on 

the file.

8.  When “Finished” appears in the display, the motor will stop and the calibration will be completed (press the Main switch   to 

exit the calibration mode).

Recommended settings for FKG instruments

Further information

Position not recommended

•  Finger position

For reprocessing or complicated treatments requiring strong pressure from the operator, in order to reduce stress at the 

intersection between the contra angle and the motor handle, place your fingers on the countra angle rather than the motor 

handle.

Recommended position

iRace

ScoutRace

Race ISO 10

BT-Apisafe

BT-Race

Race

BioRace

Shaper

Finisher

Finisher R

DR1

DR2

Op

mal Speed (rpm)

Speed*

1'000

1'000

1'000

600

800

800

800

800

600

600

1'000

600

Reverse Torque (Ncm)

Rev. Torque*

1

1

1

1.5

1.5

1.5

1

1.5

1.5

1.5

1.5

1.5

Rota

on Mode

Rot. Mode*

Torque Slow Down

Torq. SL.D.*

Linked Func

on

Linked Func

on*

Apical Reverse or Stop

Apical ACT.*

Auto Start/Stop

Auto STT/STP*

Apical Slow Down

Apical SL.D.*

Apical Torque Reduc

on

Apical TRQ.D.*

M

ai

n

 

se

n

gs

Forward (Fwd*)

Su

b

-s

e

n

gs

These func

ons are not required for FKG instruments and can therefore be deac

vated.

Se

ngs

XP-endo®

D-Race

•  Restore default memories

Restore the initial settings for the memories in the following way:

1.  Hold down the Select switch  , Plus and Minus switches 

 and then turn the unit on with the Main switch  .

2.  The “MemClear” display will appear. Press the Select switch   to restore the default memories or press the Main switch   

to cancel the operation.

3.  Wait unit the « Finished » display appears and then press the Main switch   to go to the standby display. 

It is impossible to 

restore the settings of a single memory.

  For more information, please visit  

www.fkg.ch

 and/or scan the QR code above.

FKG - 2017-10 - © FKG Dentaire SA - Pictrues of our products are not contractual.

* Abbreviation of the function according to the display of the Rooter S.