MAN-00112 R 2.0 - EasyGrid Base User Guide
5
2.3 Declaration of conformity
EU Declaration of Conformity
Manufacturer Name:
FISO Technologies Inc.
Address
:
500, Saint-Jean-Baptiste Ave., Suite 195, Quebec, QC, Canada G2E 5R9
We hereby declare under our sole responsibility that the following apparatus:
Product description:
Fiber optic thermometer
Model number(s):
EasyGrid Base
Product category:
Electrical equipment for measurement, control and laboratory use.
Complies with the essential requirements of the following applicable European Directive:
2014/30/EU Electromagnetic Compatibility (EMC) Directive,
2011/65/EU Restriction of the use of certain Hazardous Substances (RoHS) Directive
Conformity is assessed in accordance to the following standards:
EMC Standards:
EN61326-1 : 2013
Emissions/Immunity requirements for laboratory Equipment
EN60255-26 : 2013
Measuring Relays and protection equipment: electromagnetic compatibility requirements
CISPR 11:2015 A1(2016)
EN 55011:2016 A1(2017)
Measurement of conducted emission
Group 1, class A (Industrial)
CISPR 11:2015 A1(2016)
EN 55011:2016 A1(2017) Measurement of radiated emission
Group 1, class A (Industrial)
IEC 61000-4-2 : 2008
IEC 60255-22-2 : 2008
Electrostatic discharge immunity
±8kV Contact Discharge
±15kV Air Discharge
IEC 61000-4-3 : 2006
A1(2007) A2(2010)
IEC 60255-22-3 : 2007
1
Radiated electromagnetic field immunity
10V/m, 80-3000MHz
Spot frequencies : 80MHz, 160MHz, 380MHz,
450MHz, 900MHz, 1850MHz, 2150MHz: 10V/m
IEC 61000-4-4 : 2012
IEC 60255-22-4 : 2008
1
Electrical fast transient immunity (EFT)
±4kV Power Lines, ±4kV I/O
IEEE C37.90.1: 2012
Fast Transient Surge Withstand Capability
±4kV Power Lines, ±4kV I/O
IEC 61000-4-5 : 2014
IEC 60255-22-5 : 2008
1
Surge immunity
Power: ±2kV L-PE / ±1kV L-L
I/O Ports: ±2kV L-PE / ±1kV L-L
Communication Ports: ±2kV L-PE
IEC 61000-4-6 : 2013
IEC 60255-22-6 : 2001
1
Conducted disturbance induced by RF fields
10Vrms
IEC 61000-4-8 : 2009
Power Frequency Magnetic Field immunity
100A/m permanent
300A/m short duration
IEC 61000-4-9 : 2016
Pulse magnetic field immunity
1000A/m
IEC 61000-4-11 : 2004
Voltage dips, short interruptions and voltage
variation immunity
1 Cycle / 0%, 10 Cycles / 40%, 25 Cycles / 70%,
250 Cycles / 80%, 250 Cycles / 0%
IEC 60255-5 : 2000
IEC 60255-27 : 2013
Impulse voltage withstand test
±5kV
IEC 61000-4-18 : 2006
A1(2010)
IEC 60255-22-1 : 2007
1
IEEE C37.90.1 : 2012
Burst disturbance
2.5kV 2s
IEEE C37.90 : 2005
Dielectric Strength (RS-485 Circuit Only)
4.2kV
1
These standards have been replaced by IEC 60255-26:2013
2
This standard has been replaced by IEC 60255-27:2013