background image

 

www.eu.fishersci.com/fisherbrand

 

30 

h)  Press Dispense button again to enter 

Ready mode. 

 

 

 

 

 

i)  In Ready mode, press 

  and 

 

simultaneously to enter user menu. 

Press 

  to move cursor to PACK

Press 

  to reset cartridge life. Newly 

replaced cartridge shows 360 days life 

and 0 L water purified 

 

j)  Install the final filter as described in 

3.2.3. 

 

 

k)  In Ready mode, press 

  and 

 

simultaneously to enter user menu. 

Press 

  to move cursor to FILTER

press 

  to reset final filter life. Newly 

replaced final filter shows 360 days life. 

 

 

PACK: 360 DAYS      0L 

FILTER: 360 DAYS 

UV: 720DAYS      0Hr 
FLOW: 1.50 L/min 
UNIT: M

.cm 

TIME: 2014-11-27 09:21:27 
LANGUAGE/

语言:

ENGLISH 

SERIAL NO: S4PPAL3701 

 

PACK: 360 DAYS      0L 

FILTER: 360 DAYS 
UV: 720DAYS      0Hr 
FLOW: 1.50 L/min 
UNIT: M

.cm 

TIME: 2014-11-27 09:21:27 
LANGUAGE/

语言:

ENGLISH 

SERIAL NO: S4PPAL3701 

 

Содержание Accu100

Страница 1: ...Accu100 Ultrapure Water System With Remote Dispenser USER MANUAL...

Страница 2: ...LES We certify that these water systems are designed and manufactured in accordance with the following European Council directives and FCC rules The Essential Safety Requirements of EU Low Voltage Dir...

Страница 3: ...www eu fishersci com fisherbrand 3 This Manual is Applicable to the Following Accu100 Models 15409759 Accu100 UV Ultrapure Water System with Dispenser 100 240 VAC...

Страница 4: ...2 4 Installation 12 3 SYSTEM START UP AND OPERATION 17 3 1 Check List Prior to System Start Up 17 3 2 system start up and Setting 18 3 3 Routine Operation 22 3 4 Volumetric Dispensing 23 4 MAINTENANCE...

Страница 5: ...nuous flow of high quality pure water Please contact us or your local distributor if you encounter any issues during installation and use Professional engineers are fully trained to support you Safety...

Страница 6: ...sin and fine activated carbon to ensure long useful life and a low running cost A 0 2 m final filter or an ultrafiltration filter optional removes any remaining contaminants A dual UV lamp to kill bac...

Страница 7: ...ting Voltage 100 240 V Max Power 55 W System Dimensions Width height x depth 8 17 14 in 21 44 36 cm Dispenser Dimensions Diameter height 9 29 in 22 74 cm Water Production Rate 1 1 5 L min Resistivity...

Страница 8: ...The Accu100 system comes with the following items in the package The following items are packed separately These items need to be installed at initial set up 1 Main system 2 User Manual and Quality C...

Страница 9: ...out 2 3 2 Remote Dispenser Diagram Feed Water inlet In from dispenser Purification cartridge Cartridge lock bar Out to dispenser Cable port Power Switch In from system Out to system UP outlet LCD disp...

Страница 10: ...r recirculation Press 1 and 2 simultaneously to enter user menu 2 In Ready mode press 2 to start volumetric dispensing setting Press 1 and 2 simultaneously to enter user menu 3 In Ready mode press 3 t...

Страница 11: ...or the Accu100 and Accu100 UV models are described below Each component is explained in the legend C Booster Pump M AccuQuatro U Pack E1 Feed Water solenoid valve I Resistivity Cell E2 Recirculation s...

Страница 12: ...empt to pull the stopper out from the port with force without unlocking the collet first as this will damage the quick connector rendering it useless Please follow the instruction below to remove the...

Страница 13: ...g all the way in for about 20 mm 2 Gently pull the tubing outwards to check if the connection is secure 2 4 1 3 Connect the other end of the tubing to a feed water tank ensure the tubing inlet is imme...

Страница 14: ...he cartridge from its packaging 2 Remove blue caps on the cartridge inlet outlet 2 4 3 2 Open the lock bar at the front of the system 2 4 3 3 Remove the two blue stoppers at the water inlet outlet ins...

Страница 15: ...inet until it is seated fully inside the cabinet 2 4 3 6 Lock the front lock bar by inserting the flexible buckle to the groove If the bar cannot be locked reseat the cartridge all the way in and then...

Страница 16: ...ng tags IN and TO POU Connect the other marked tubing to the inlet FROM POU on the system and the other end Out to system to the Dispenser Tubing is marked with corresponding tags OUT and FROM POU Con...

Страница 17: ...m start up use the table below to make sure all parts have been installed and connected and quality of feed water meets minimal requirements Check List Results Feed Water quality purified water Water...

Страница 18: ...n system enter Ready mode automatically In Ready mode Press Dispense button to dispense ultrapure water system enter Product mode In Product mode Press Dispense button again to stop dispensing System...

Страница 19: ...ry bag insert one end to the adapter and the other end into the drain 3 2 2 3 1 Press Dispense button to dispense water 2 Gas inside the system will be flushed out of the system 3 2 2 4 1 Wait till no...

Страница 20: ...r outlet till finger tight no leaking at dispensing DO NOT over tight 3 2 3 3 Loosen the knob on the top of the final filter for degassing Press Dispense button to drive air out of the final filter th...

Страница 21: ...ter Life In user menu press to move cursor to FILTER and press to reset final filter life Newly replaced final filter shows 360 days life 3 2 6 Resetting UV Lamp Life In user menu press to move cursor...

Страница 22: ...o enter recirculation mode wait until resistivity reaches 18 2 M cm c Press the Dispense button to dispense water d Press the Dispense button again to stop dispensing system enter Recirculation mode a...

Страница 23: ...spensing a In Ready mode press to enter volumetric dispensing mode b Press and to set the desired amount of water c Press to dispense water d System return to Recirculation after dispensing Volumetric...

Страница 24: ...shows 360 days life Reset Final Filter Life everytime the final filter is replaced 4 1 3 Resetting UV Lamp Life In user menu press to move cursor to UV and press to reset UL lamp life Newly replaced U...

Страница 25: ...Press to move cursor to TIME Press to move cursor to corresponding number and press to increase numbers 4 1 7 Setting Language Press to move cursor to LANUGAGE Press or to switch language PACK 360 DA...

Страница 26: ...s preset the consumables service life time countdown U pack cartridges have reached their expected service life and need to be replaced EXCH FINAL FILTER Yellow System has preset the consumables servi...

Страница 27: ...bles Replacement Frequency Performance Indicator 185 254 nm UV lamp 24 months Increase in TOC Final Filter Remove bacteria and particles 12 months Reduce in flow rate When needed Flow rate is less tha...

Страница 28: ...Switch System to ready mode b Screw off the old final filter and screw on the adapter to water outlet Connect the 1 4 inch PE tubing to adapter and leave another side into drain c Press to enter Pres...

Страница 29: ...isherbrand 29 e Replace AccuQuatro U Pack as described in 2 4 3 f Press button to put system into recirculation check if there is any leakage reinstall the cartridge if leaking g Degassing the System...

Страница 30: ...he final filter as described in 3 2 3 k In Ready mode press and simultaneously to enter user menu Press to move cursor to FILTER press to reset final filter life Newly replaced final filter shows 360...

Страница 31: ...t be on at ALL TIMEs when a UV lamp is installed Caution Keep the UV lamp straight in and out of the stainless steel chamber during its installation to avoid any action that could cause the lamp to br...

Страница 32: ...ove the old UV lamp out f Wear gloves included in the new UV lamp package Avoid direct skin contact with the quartz glass of the UV lamp Carefully insert the new UV lamp into its chamber When about 2...

Страница 33: ...he system back cover and tighten the screws i In Ready mode press and simultaneously to enter user menu Press to move cursor to UV press to reset UV lamp life Newly replaced UV lamp shows 720 days lif...

Страница 34: ...y broken Check the power supply Check the main fuse Contact a ULS agent No water Gas in the system Pump failure Degassing the system Contact a ULS agent Low water flow rate Final filter clogged System...

Страница 35: ...uQuatro U Pack Cartridge RO Feed 15489709 AccuQuatro U Pack Cartridge kit with a 0 2 m Capsule Filter RO Feed 15499709 AccuQuatro U Pack Cartridge DI Feed 15409719 AccuQuatro U Pack Cartridge kit with...

Страница 36: ...nce Spare Parts Contact Fisher Scientific for ordering information Main control panel 100 240V 50 Hz Recirculation Pump Resistivity Sensor Commonly Used Connector Package UV lamp ballast Power Supply...

Страница 37: ...nd workmanship when used in compliance with instructions and operating conditions specified in this manual Thermo Fisher Scientific warrants the system for 24 months from the earlier of 1 The date of...

Отзывы: