| Operation
28
|
50152755
Fisherbrand
Chapter 6
4. After the preset temperature is reached load the work space with samples.
Hot surfaces
The screen of the glass door, the interior panel of the outer door as well as the
surfaces of the shelving and the work space become hot while the incubator is
running through its heating cycles and need some time to cool down.
When removing samples from a running or recently completed heating cycle, always
wear safety gloves and other appropriate personal protection equipment to avoid
burns on hot surfaces!
Risk of overloading
Overloading may damage the shelves or cause the shelves and/or the incubator to tilt
when the shelves are being drawn out, ultimately destroying the samples. To avoid
overloading the incubator or its shelving be sure to observe the sample weight limits
specified in Chapter 13, “Technical Data.”
NOTE
Proper loading
To ensure sufficient air circulation and uniform heating of the samples, do not use more
than 70% of the maximum surface area of the work space. Bulky objects in the work
space that dissipate heat may impair heat distribution.
NOTE
Door opening
Keep the door open-times during operation always as short as possible. The longer the
door is open, the higher can be a possible overshooting of the temperature.
Содержание 117L
Страница 1: ......
Страница 6: ...2 part number Fisherbrand...
Страница 12: ...Safety Notes 6 50152755 Fisherbrand Chapter 1...
Страница 26: ...Product Description 20 50152755 Fisherbrand Chapter 4...
Страница 32: ...Start up 26 50152755 Fisherbrand Chapter 5...
Страница 38: ...Handling and Control 32 50152755 Fisherbrand Chapter 7...
Страница 40: ...Shut down 34 50152755 Fisherbrand Chapter 8...
Страница 50: ...Disposal 44 50152755 Fisherbrand Chapter 11...
Страница 52: ...Error Codes 46 50152755 Fisherbrand Chapter 12...
Страница 58: ...Accessories 52 50152755 Fisherbrand Chapter 14...
Страница 60: ...Device Log 54 50152755 Fisherbrand Chapter 15...