Safety Notes  |  

Chapter 1

Intended Purpose of the Incubator

Correct Use

Fisherbrand incubators are laboratory devices for preparing and cultivating cell and tissue 
cultures. The devices employ precision temperature control for simulating the specific 
physiological ambient conditions for these cultures.

Fisherbrand incubators are for professional use only.

Incorrect Use

To avoid the risk of explosion do not load the incubator with tissue, material, or liquids that:

are easily flammable or explosive,

release vapor or dust that forms combustible or explosive mixtures when exposed to air,

release poisons,

do not pour any liquids on the bottom of the interior surface or into a collecting basin inside 
the unit.

release dust

exhibit exothermic reactions

are pyrotechnical substances

refrain also from pouring any liquids onto the internal base plate or inserting bowls filled with 
liquids into the sample compartment. 

Standards and Directives

The incubator complies with the following standards and guidelines:

IEC/CAN/CSA-C22.2/UL 61010 - 1

IEC/CAN/CSA-C22.2 61010 - 2 - 010

China EEP Hazardous Substances Information
http://www.thermofisher.com/us/en/home/technical-resources/rohs-certificates.html.

Additionally, the incubator is in compliance with many other international standards, regulations 
and directives not listed here. Should you have any questions regarding compliance with national 
standards, regulations and directives applicable for your country, please contact your Fisherbrand 
sales organization.

Содержание 117L

Страница 1: ......

Страница 2: ...ylvania 15275 United States Thermo Fisher Scientific provides this document to its customers with a product purchase to use in the product operation This document is copyright protected This document...

Страница 3: ...rs 8 Chapter 4 Installation 9 Ambient Conditions 9 Location Requirements 9 Intermediate Storage 10 Room Ventilation 10 Incubator Dimensions and Clearances 11 Transport 12 Installing the Anti tilt Anch...

Страница 4: ...Down 33 Chapter 10 Cleaning and Disinfection 35 Cleaning 35 Cleaning Exterior Surfaces 35 Wipe Spray Disinfection 35 Preparing the Manual Wipe Spray Disinfection 37 Predisinfection 37 Cleaning 37 Fina...

Страница 5: ...and Basic 60L 100L 180L Gravity Incubator Rear View 16 Figure 4 3 Sensor System 17 Figure 4 4 Basic Gravity Incubator Shelf System 19 Figure 5 1 Sliding the Retaining Spring into the Support Rail 21 F...

Страница 6: ...2 part number Fisherbrand...

Страница 7: ...sued by the operator shall be used to create written procedures targeted at personnel working with the subject matter device detailing the decontamination measures to be employed for the incubator and...

Страница 8: ...s Check the door seal every six months for proper sealing performance and possible damage Do not process any samples containing hazardous chemical substances that may be released into the ambient air...

Страница 9: ...ndicates a hazardous situation which if not avoided will result in death or serious injuries Indicates a hazardous situation which if not avoided could result in death or serious injuries Indicates a...

Страница 10: ...4 50152755 Fisherbrand Chapter 1 Symbols on the Incubator Electric shock Hot surfaces Fire hazard Explosion hazard Suffocation hazard Danger of tipping Observe operating instructions Mark of conformi...

Страница 11: ...m of the interior surface or into a collecting basin inside the unit release dust exhibit exothermic reactions are pyrotechnical substances refrain also from pouring any liquids onto the internal base...

Страница 12: ...Safety Notes 6 50152755 Fisherbrand Chapter 1...

Страница 13: ...pection After the incubator has been delivered check the delivery immediately for completeness possible damage If components are missing or damage is found on the incubator or the packaging in particu...

Страница 14: ...s Quantity of components supplied pieces Gravity Convection Incubators Perforated shelves 2 Support rail for shelf table top incubators 4 Shelf support 4 Power cord 1 Clip springs for table top incuba...

Страница 15: ...accessories particularly if two devices are stacked The electrical circuitry of the incubator has been designed for an operating height of up to 2000 m above sea level Relative humidity up to 80 maxim...

Страница 16: ...evices are to be placed in the same room additional ventilation may have to be provided as necessary To avoid any impact of the heat dissipated by the incubator on the ambient climate the room must be...

Страница 17: ...0L Gravity Incubator 120 V 640 25 2 738 29 1 920 36 2 650 25 6 Fisherbrand 75L Incubator 230 V 530 20 8 565 25 2 720 28 3 540 21 3 Fisherbrand 117L Incubator 230 V 640 25 2 565 25 2 820 32 3 650 25 6...

Страница 18: ...re To avoid injury through physical strain such as strain traumata and slipped discs do not attempt to lift the incubator alone To avoid injury through dropped loads be sure to wear Personal Protectio...

Страница 19: ...the bracket screws Use the preferred position if possible 5 Fix the anti tilt anchor with the bracket side down to the unit 6 Position the unit with the anti tilt anchor to in an angle of approx 90 20...

Страница 20: ...specified stacking height Risk of tipping and dropping of stacked devices You should be aware at all times that stacked devices do not form a stable unit even when the stacking pads and frames are co...

Страница 21: ...to 75 C 167 F two perforated shelves access port for table top units The individual features of gravity convection incubators are shown in the figures below Figure 4 1 Fisherbrand Basic 60L 100L 180L...

Страница 22: ...eatures an overheat protection cut out feature that shuts down the incubator completely when excessive temperatures occur in the workspace dual fuses rated at 16 amperes Work Space Atmosphere To ensur...

Страница 23: ...ed to the AC supply mains via the socket item 4 in Figure 4 4 which accepts a power cord with an IEC standard plug Fuses Two 16 A slow blow fuses T 16A H 250 V mounted on the incubator s main electron...

Страница 24: ...hoses or additional sensor leads to be routed into the work space of the incubator The small access port 1 is secured with a membrane Cut off before use The main access port 2 has a diameter of 42 mm...

Страница 25: ...30 mm allowing the shelf support brackets 8 to be inserted without any room for error yet in a very flexible way to accommodate any required height of sample container The shelves 2 have an integrated...

Страница 26: ...Product Description 20 50152755 Fisherbrand Chapter 4...

Страница 27: ...the rails the perforated shelves can be simply pushed onto their support hooks to complete the installation Initial Installation 1 Peel off the protective foil from the support rails 2 Push the retain...

Страница 28: ...and 5 act as lateral guides for the support rails while the embossings at 1 and 6 secure the support rails in place For the support rails to install correctly the retaining spring 3 must be facing upw...

Страница 29: ...Installation Installing the Perforated Shelves 1 Rear Pull out Stop 2 Anti tilt anchor 3 Front Pull out Stop 4 Shelf Figure 5 4 Installing the Perforated Shelves 1 Push the shelf 4 onto the shelf sup...

Страница 30: ...nt of the incubator If the voltage V and current A ratings given are not as required do not connect the incubator to the power source 2 Connect the IEC connector to the socket at the rear of the incub...

Страница 31: ...1 paragraph 6 2 2 The user must ensure that the device is operated on a mains that complies with this requirement Where necessary it is recommended to have the system impedance confirmed by the power...

Страница 32: ...Start up 26 50152755 Fisherbrand Chapter 5...

Страница 33: ...alled safely Disinfecting the Incubator s Work Space Disinfect the work space according to the operator specified hygiene guidelines Starting Operation 1 Turn the incubator on using the control panel...

Страница 34: ...faces Risk of overloading Overloading may damage the shelves or cause the shelves and or the incubator to tilt when the shelves are being drawn out ultimately destroying the samples To avoid overloadi...

Страница 35: ...ontrol functions and adjustments of the incubator The graphic below shows the control panel with all of its visualization elements and controls Figure 7 1 Control Panel for Fisherbrand Basic Gravity I...

Страница 36: ...use only Down button Up button Function button scroll button Upper Display Lower Display Description PV Value Active SP Value PV measuring value and target value of selected SP setpoint SP Value SP Se...

Страница 37: ...set the required value in 0 1 C steps Acknowledge the entry with or wait approx 2 min for automatically storage of the value The new temperature is adjusted For example The controller adds the adjuste...

Страница 38: ...Handling and Control 32 50152755 Fisherbrand Chapter 7...

Страница 39: ...he control panel 4 Unplug the power cord and secure it against accidental reconnection 5 Until the incubator is shut down the work space must be continuously ventilated Leave the glass door and the ou...

Страница 40: ...Shut down 34 50152755 Fisherbrand Chapter 8...

Страница 41: ...n Incompatible cleaners Some incubator components are made of plastics Solvents may attack plastics Strong acids or alkaline solutions may cause embrittlement of plastics Please consult the manufactur...

Страница 42: ...adequately ventilated rooms After the disinfectant has been allowed to react wipe the cleaned incubator components thoroughly dry Observe safety regulations to avoid fire and or explosion hazard cause...

Страница 43: ...es 3 to 5 times with autoclaved water to completely remove and cleaning agent residues 4 After this wipe the cleansed surfaces and internals dry with a soft sterile cloth Health hazard The surfaces of...

Страница 44: ...amiliar with the internationally recognized Laboratory Biosafety Manual from the World Health Organization WHO and with national pertinent recommendations These manuals recommendations provide informa...

Страница 45: ...and proper seating of the seal Perform functional check of the control panel and of the incubator s built in controller Perform electrical safety check in accordance with the relevant national regula...

Страница 46: ...monthly service Perform the comparative temperature measurement outlined in section 7 3 Annual Service Have the incubator inspected and services by a Fisherbrand authorized Technical Service agent NOT...

Страница 47: ...Gently press the seal into the slot working around the circumference of the door In corner areas in particular ensure that the seal lip is installed without any wrinkles and that the seal is not stret...

Страница 48: ...ch may have caused contamination of the incubator and its components Prior to return shipment it is therefore mandatory that all incubator components be properly decontaminated Clean the incubator com...

Страница 49: ...lation components Glass wool Printed circuit boards Coated electrical components contain various plas tics materials Components mounted on circuit boards containing epoxy resin bonder Plastic componen...

Страница 50: ...Disposal 44 50152755 Fisherbrand Chapter 11...

Страница 51: ...perature is 2 C above setpoint The controller is operating the incubator back to set temperature If an alarm is indicated AlSt 1 is dis played after pressing twice the Enter but ton Sensor Break OPEN...

Страница 52: ...Error Codes 46 50152755 Fisherbrand Chapter 12...

Страница 53: ...ature deviation from set value at 37 C 98 6 F spatial typical max K 0 8 0 9 0 8 0 9 0 8 0 9 Temperature deviation from set value at 37 C 98 6 F over time K 0 3 0 3 0 3 Heat up time work space unoccupi...

Страница 54: ...r F r H 80 non condensing Min storage temperature C F 20 68 Max storage temperature C F 60 140 Max humidity in storage non condensing r F r H 90 non condensing Post transport acclimation time h 2 Nois...

Страница 55: ...inal temperature min 80 80 80 Recovery time work space unoccupied door open for 30 s to 98 starting temperature typical max min 8 10 10 12 12 14 Heat dissipation to environment at set temperature of 3...

Страница 56: ...ture C F 20 68 Max storage temperature C F 60 140 Max humidity in storage non condensing r F r H 90 non condensing Post transport acclimation time h 2 Noise level dB A no Inherent noise Pollution degr...

Страница 57: ...ed shelf for 100L 117L incubators including 2 shelf supports 150145854 Stainless steel perforated shelf for 180L 194L incubators including 2 shelf supports 150145846 Wire mesh shelf for 60L 75L incuba...

Страница 58: ...Accessories 52 50152755 Fisherbrand Chapter 14...

Страница 59: ...Fisherbrand 50152755 53 Device Log Incubator type Part number Serial number Service number Location Operator s note Work carried out Notes Date Signature...

Страница 60: ...Device Log 54 50152755 Fisherbrand Chapter 15...

Страница 61: ...22 95 59 59 Sweden 46 31 352 32 00 Belgium 32 0 56 260 260 Ireland 353 0 1 885 5854 France 33 0 3 88 67 14 14 Portugal 351 21 425 33 50 Switzerland 41 0 56 618 41 11 Denmark 45 70 27 99 20 Italy 39 0...

Отзывы: