B. 

To adapt a video camera an adapter tube is needed. This adapter tube includes 0.5X lenses that 
correct the parfocality of images, from both binocular and vertical ports, shown on the TV monitor. 
This is providing that parfocality has been correctly adjusted according to the steps followed in point 
“D. Changing magnification” in this manual.  
The adapter tube is provided with a “C” thread, and a “CS” ring which can adapt different kinds of 
video camera.  
To connect the camera to the microscope, screw the adapter tube to the video camera.  

 

1.  Loosen the knurled screw (Fig.2) on the side of the vertical port of the head sufficiently enough 

to remove the protective cover.  

 

2.  Insert the adapter tube with the camera already mounted on the vertical port, as shown in figure 

4. If it does not insert easily, unscrew the knurled screw until the adapter tube fits perfectly.  

 

3.  Re-tighten the knurled screw firmly, so that the camera is secure.  

 

4.  Operate the camera according to manufacturers instructions.  

 

If the image on the monitor appears out of focus when the objective is changed, it is possible that the 
CS mount is responsible. Place, or remove it, according to the procedure to obtain parfocality.   

 
 

Maintenance 

 

WARNING:

 FOR YOUR OWN SAFETY SWITCH OFF AND DISCONNECT THE MICROSCOPE FROM 

ANY ELECTRICAL SOURCE BEFORE ATTEMPTING ANY MAINTENANCE PROCEDURE TO AVOID THE 
RISK OF ELECTROCUTION.  
CONSULT YOUR DISTRIBUTOR IF ANY REPAIR OR MAINTENANCE PROCEDURE IS REQUIRED TO 
YOUR MICROSCOPE THAT DOES NOT APPEAR IN THIS INSTRUCTION MANUAL. 
 
A. Optical 

maintenance. 

 

Do not attempt to disassemble any optical component. For any repair work not specified in this manual, 
consult the technical service responsible in your area.  

 

Before cleaning the lens surface, remove dust with a brush specifically for lenses, or with low pressure 
compressed air, found in any photography shop.  

 

1. Cleaning 

the 

eyepiece 

 

a.  Do not remove the eyepiece (1) from the eyepiece tube (3). 

 

b.  Clean the external surface dampening the lens with breath. 

 

c.  Afterwards, dry the lens with special lens paper. Dry in circular movements from the centre o 

the lens, outwards. Do not wipe the lens when already dry, as they scratch easily.   

 

2. Cleaning 

the 

objectives. 

 

a.  Do not remove objectives from the microscope. 

 

b.  Only clean the surface area. Use a soft cotton cloth dampened slightly with Xylene. Dry the 

lens afterwards with the same cloth.  

 
 
 
 
 
 
 
 

-7- 

Содержание SMZ-140-N2GG

Страница 1: ...Instruction Manual SMZ Series Copyright 02 02 Motic Microscopes European Division ...

Страница 2: ... Support collar screw 18 Focus knob 1 Eyepiece 14 Head 2 Diopter adjuster 3 Eyepiece tubes 5 Head holder locking screw 17 Head holder 15 Power cord for incident illumination 4 Prism housing 13 Base 8 Incident light adjustment screw 6 Objectives 7 Incident illumination 9 On Off switch 12 Stage plate locking screw 11 Stage 16 Zoom knob SMZ 140 N2GG 2 ...

Страница 3: ...nents of the stereomicroscope have been carefully packed to make sure they reach you in perfect condition We recommend that you do not discard any packing containers in case you need to return the microscope or store it for long periods of time or should it become necessary to transport it to a technical service for any repair or maintenance procedure The box should contain the following component...

Страница 4: ...ings etc The black and white stage is used for non transparent objects or for dissection For best contrast choose the side of the stage to use Warning Transmitted illumination can ONLY be used with the frosted glass stage THE HEAT GENERATED BY THE TRANSMITTED ILLUMINATION CAN MELT OR DAMAGE THE BLACK AND WHITE STAGE SUCH DAMAGE WOULD NOT BE INCLUDED UNDER WARRANTEE 1 Changing the stage a Loosen th...

Страница 5: ...entre of the stage 11 3 Turn focussing knobs 18 to mid focus range 4 The head holder 17 is mounted on a column 21 on which it can be moved up or down depending on the size of the object to be focussed on a Support head holder 17 with one hand without touching any lens and with the other loosen the screw 20 on the support collar 19 The head holder can then be slid to the base 13 b Without letting g...

Страница 6: ...extended position the image is transmitted to the vertical port but the image cannot be seen by the right eyepiece A To adapt a photographic camera an adapter tube is needed This tube includes 2X lenses to balance the correct parfocality between the images received by the binocular and vertical ports This balance can be achieved providing that the parfocality has been adjusted correctly as describ...

Страница 7: ...that the CS mount is responsible Place or remove it according to the procedure to obtain parfocality Maintenance WARNING FOR YOUR OWN SAFETY SWITCH OFF AND DISCONNECT THE MICROSCOPE FROM ANY ELECTRICAL SOURCE BEFORE ATTEMPTING ANY MAINTENANCE PROCEDURE TO AVOID THE RISK OF ELECTROCUTION CONSULT YOUR DISTRIBUTOR IF ANY REPAIR OR MAINTENANCE PROCEDURE IS REQUIRED TO YOUR MICROSCOPE THAT DOES NOT APP...

Страница 8: ...t illumination bulb a Unscrew the illuminator protector tube 7 turning it anti clockwise and removing the tube from that of the lenses b With a cloth carefully pull out the bulb to disconnect it from the socket g If the bulb is accidentally touched with bare hands it must be cleaned as this could affect the transmission of light and life span of the bulb c Replace illuminator lens tube turning in ...

Страница 9: ...ely as possible without the head sliding down with its own weight The tension adjustment collar for focussing Fig 1 is situated between the focussing knob 18 and the head holder 17 a Loosen the screw located in the collar hole with the 2mm hexagonal key b To increase the tension turn the collar anti clockwise to decrease it turn the collar clockwise c Re tighten the hexagonal key 10 ...

Страница 10: ...Fuse cover badly closed Bad connection with power source Insert bulb correctly Replace bulb Close correctly Have repaired by a qualified specialised technician Fuse blown quickly Wrong fuse Replace with appropriate fuse Fuse blown instantly Short circuit Have repaired by a qualified specialised technician IMAGE QUALITY PROBLEM CAUSE SOLUTION Poor resolution Eyepieces dirty Objectives dirty Clean e...

Страница 11: ...ails of the required revision Warrantee All MOTIC microscopes are warranted against any manufacturing defect for a 5 year period Damage occurring by any unauthorised repair work or occurring through misuse or modification of the microscope will not be included under the conditions of the warrantee Bulbs and fuses are not under warrantee The warrantee service is provided by MOTIC or its authorised ...