Operator‘s Manual FERITSCOPE
®
FMP30
Page 191
Malfunctions and Messages
11.2 Messages on the Display
The error messages (
E***
) and warning messages (
W***
) that may occur
during instrument operation are contained in the overview on the following
pages.
Malfunction/Message Cause
Correction
start
on
Page
Reading cannot be dis-
played (because the val-
ue is greater than 9999
or smaller than
- 9999)
Cause: Measurement
carried out incorrectly.
Perform the measure-
ment correctly (e.g., do
not allow the probe to
hover over the speci-
men before or after the
measurement; observe
the minimum lift-off dis-
tance!)
Cause: Eddy current
probe that is currently
normalized on a non-fer-
romagnetic base materi-
al has previously been
normalized on a ferro-
magnetic base material.
Continue normaliza-
tion.
Instrument is not ready
to make measurements.
Cause: Re-initialization
was carried out.
Application was delet-
ed and then switched
off.
Use
ZERO
to start and
perform the normaliza-
tion.
Histogram does not print
because fewer than 30
readings are stored in
the current Application.
Take more than 30
measurements.
- - - -
E
000
Not enough
measurements
!
Содержание FERITSCOPE FMP30
Страница 1: ...FERITSCOPE FMP30 Operator s Manual Coating Thickness Material Analysis Microhardness Material Testing ...
Страница 63: ...Page 58 Operator s Manual FERITSCOPE FMP30 Applications Fig 5 3 Example for linking of applications ...
Страница 235: ...Page 230 Operator s Manual FERITSCOPE FMP30 Glossary ...
Страница 240: ...Operator s Manual FERITSCOPE FMP30 Page 235 X Xn 79 ...
Страница 241: ...Coating Thickness Material Analysis Microhardness Material Testing ...