Page 34
Operator‘s Manual FERITSCOPE
®
FMP30
Probe Handling
4
Probe Handling
4.1
Handling During Measurements
Always hold the probe at its grip sleeve (right
figure).
Always place the probe gently and at a right
angle on the specimen surface.
Slide the grip sleeve to the specimen surface
such that the sleeve rests on the specimen (center
and bottom Figure, right).
With the default setting, a beep will signal the
measurement capture.
Lift the probe off the specimen before making the
next measurement.
Avoid hard impacts.
Do not allow the probe to hover directly above the surface. Doing so
will lead to erroneous readings.
Do not bend the probe connector cable! Doing so can lead to broken
wires.
Example:
Probe
FGAB1.3
Grip
Specimen
Содержание FERITSCOPE FMP30
Страница 1: ...FERITSCOPE FMP30 Operator s Manual Coating Thickness Material Analysis Microhardness Material Testing ...
Страница 63: ...Page 58 Operator s Manual FERITSCOPE FMP30 Applications Fig 5 3 Example for linking of applications ...
Страница 235: ...Page 230 Operator s Manual FERITSCOPE FMP30 Glossary ...
Страница 240: ...Operator s Manual FERITSCOPE FMP30 Page 235 X Xn 79 ...
Страница 241: ...Coating Thickness Material Analysis Microhardness Material Testing ...