background image

Isotemp Plus

General Purpose Laboratory Freezers

Catalog Numbers:

13-986-142/142R

13-986-143/143R

13-986-144/144R

13-986-147/147R

104215•

 Rev. A

Содержание Isotemp Plus 13-986-142

Страница 1: ...Isotemp Plus General Purpose Laboratory Freezers Catalog Numbers 13 986 142 142R 13 986 143 143R 13 986 144 144R 13 986 147 147R 104215 Rev A...

Страница 2: ...Casters 5 Shelves 6 Electrical Connection 6 Operation 7 Control Layout 7 Setting the Temperature Control Setpoint 7 Units Select 8 Calibration Offset 8 Hold Off Time 9 Temperature Digression 9 Automat...

Страница 3: ...s intended application alterations or modifications will void the Warranty This product is not intended nor can it be used as a ster ile or patient connected device In addition this appara tus is not...

Страница 4: ...such external evidence of loss or damage may result in the carrier s refusing to honor your damage claim The carri er will supply the form required to file such a claim Concealed Loss or Damage Conce...

Страница 5: ...it After installation confirm that the unit was installed per these instructions that the required information was completed and that the form was faxed to the specified number Installing and Leveling...

Страница 6: ...e customer should have the wall receptacle and circuit checked by a qualified electrician to verify the receptacle is properly grounded and is connected to the proper power supply The following list g...

Страница 7: ...icated by the C or F LED located just to the right of the temperature display The refrigerator is factory set at 15 C When the unit is first energized there is a delay time before the compressor will...

Страница 8: ...nd control point to a ref erence standard To set the temperature offset press and hold the Menu keypad for 5 seconds The display will flash oS offset followed by the last temperature off set value The...

Страница 9: ...sing the Increase keypad the display will show the warmest temperature recorded since the unit was ener gized or the digression values were reset Pressing the Decrease keypad will display the coldest...

Страница 10: ...t and lukewarm water or a solu tion of bicarbonate of soda 1 tablespoon per gallon of water is recommended for cleaning the interior and exte rior of the cabinet All surfaces should be rinsed and thor...

Страница 11: ...ondensate evaporator pan located behind and below the condenser fan must be cleaned periodically to prevent foul odors and to operate efficiently Vacuum clean if dry or sponge clean with soap and wate...

Страница 12: ...ot A clicking sound occurring about every 30 seconds indicates this protector is working Insufficient cooling Set temp is too high Reduce temperature setting verify Cool LED not lit Cool LED is on Con...

Страница 13: ...t 1 800 438 4851 Item Part No Shelves SPN13986210E Compressor Relay K1 SPN88616 Temperature Controller SPN104161 International Version Controller SPN105402 Chart Recorder SPN104114 Chart Recorder Inst...

Страница 14: ...K DEFROST SWITCH 1 BLK BLK YEL BLU L1 Compressor Relay K1 COMPRESSOR MOTOR CONDENSER FAN S K1 BRN DEFROST TIMER CUT OUT COIL ORN RED DEFROST LIMIT CONTROL WHT DEFROST HEATER CONDENSATE DRAIN HEATER OR...

Страница 15: ...sor Relay Cool Relay R1 CONDENSATE DRAIN HEATER COMPRESSOR MOTOR CONDENSER FAN S DEFROST HEATER WHT BRN PNK R1 DEFROST TIMER CUT OUT COIL DEFROST SWITCH 1 ORN ORN DEFROST LIMIT CONTROL RED BLK YEL YEL...

Страница 16: ...m the respective manufacturer whatever adjustment is reasonable in the light of the manufacturer s own warranty b The Company shall be released from all obligations under its warranty in the event rep...