Setting the Temperature
To change the set temperature, press and release the
Menu
keypad once. The display will flash SP and the
Mode LED will be illuminated. The last set temperature
will then be shown in the display. To change the tempera-
ture, press the
UP
or
DOWN
arrow key. The adjustable
temperature range is 0 to -30°C (32 to -22°F). When the
desired set temperature is displayed, press the
Menu
key-
pad to enter the set temperature and activate the
Units
Select
menu.
Units Select
The second menu selects the units. The freezer control
displays temperature in °C or °F. The factory setting is
°C. Starting from the temperature display mode (Mode
LED off), press the
Menu
key twice. The current temper-
ature units are displayed. To switch between units, press
the UP or DOWN arrow key to change to the desired
units, then press the
Menu
key to select. The desired
unit LED will be displayed on the right of the temperature
display.
Calibration Offset
In the event the freezer needs calibrated, a simple routine
is available to adjust the display and control point to a ref-
erence standard. To set the temperature offset, press
and hold the
Menu
keypad for 5 seconds. The display
will flash oS (offset) followed by the last temperature off-
set value. The factory setting is 0. To change the offset
value, press the UP or DOWN arrow key, then press the
Menu
key. The value shown in the display will be added
to the previous temperature reading. The maximum off-
set value is ±5°C (±9°F).
For Example:
The freezer has been operating at -15°C for 30 minutes.
The display indicates -15°C but a reference thermometer
in the chamber indicates -17°C. The operator presses
and holds the
Menu
keypad for 5 seconds and changes
the offset value from 0 to -2. The
Menu
key is pressed
until the
Mode
LED goes out. Now the display indicates a