Operation    

Continuous Operation

1.  Press the 

MINUS / PLUS

 buttons 

 until 

hd

 is shown.

TIME

2.  During continuous operation, the centrifuge will continue running until you stop it 

manually with 

STOP

.

CAUTION 

Please note that the lifetime of rotor tubes is limited, in particular if the tubes are made 

of plastics materials. 

Continuous operation (extended use) may cause damage to them.

3. 5. 5.   Pre-selecting Temperature

You can determine the sample temperature in °C. Proceed as follows:

1.  Enter the desired value by pressing the 

MINUS / PLUS

 buttons 

 under the right 

display repeatedly, until the desired value shows. You increase or decrease the 
temperature in steps of 1 °C.

2.  Press 

START

 to confirm the pre-selected value.

If you do not press any button, the display flashes for a few seconds. The new 
preselected value is now stored and the display shows the actual value.

TEMP

COOL

3.  Close the centrifuge lid.

4.  Restart the centrifuge.

The refrigeration starts operating if the pre-selected temperature is below the current 
temperature of the rotor chamber.

Содержание Fisherbrand accuSpin Micro Series

Страница 1: ...Fisherbrand accuSpin Micro Series Instruction Manual 50165241 b 2022 03...

Страница 2: ...Micro 17 12 Fisherbrand accuSpin Micro 17R 13 Fisherbrand accuSpin Micro 21 14 Fisherbrand accuSpin Micro 21R 15 1 3 1 Directives and Standards 16 1 4 Mains Supply 18 1 5 Refrigerants 19 1 6 Rotor Spe...

Страница 3: ...CF Value Explained 46 3 5 Entering Centrifugation Parameters 47 3 5 1 Selecting Speed or RCF Value 47 3 5 2 Pre selecting Speed 47 3 5 3 Pre selecting the RCF Value 48 3 5 4 Pre selecting Run Time 48...

Страница 4: ...ection 58 4 5 Decontamination 58 4 6 Autoclaving 59 4 7 Service 59 4 8 Lifetime 60 4 9 Shipping 60 4 10 Storage 60 4 11 Disposal 61 5 Troubleshooting 62 5 1 Mechanical Emergency Lid Release 62 5 2 Ice...

Страница 5: ...aboratories by trained personal Signal Words and Symbols Signal Word Degree of Hazard WARNING Indicates a hazardous situation that if not avoided could result in death or serious injury CAUTION Indica...

Страница 6: ...y instructions The centrifuge is to be used for its intended use only Improper use can cause damages contamination and injuries with fatal consequences The centrifuge must be operated by trained perso...

Страница 7: ...and any local regulations When centrifuging microbiological samples from the Risk Group II according to the Laboratory Biosafety Manual of the World Health Organization WHO aerosol tight biological s...

Страница 8: ...e occurs Do not open the centrifuge before the rotor has stopped spinning Do not touch a spinning rotor Open the centrifuge only if the rotor has stopped spinning Never use your hands or tools to stop...

Страница 9: ...it is running The centrifuge housing is not to be opened by the operator CAUTION Due to air friction sample integrity may be affected The temperature of the rotor may rise significantly while the cen...

Страница 10: ...main switch Pull out the power supply plug In an emergency disconnect the power supply Make sure that the main switch and power supply plug can be accessed easily when setting up the centrifuge The g...

Страница 11: ...rbrand accuSpin Micro 21R 120 V 10 60 Hz 75003248 Fisherbrand accuSpin Micro 21 230 V 10 50 60 Hz 75003249 Fisherbrand accuSpin Micro 21R 230 V 10 50 60 Hz Table 1 List of Centrifuges 1 2 List of Roto...

Страница 12: ...at Dissipation 0 15 kW h Environmental Conditions For Storage and Shipping Temperature 2 C to 50 C For Operation Use in interior spaces Altitudes of up to 3000 m above sea level Temperature 2 C to 40...

Страница 13: ...ion 0 25 kW h Environmental Conditions For Storage and Shipping Temperature 2 C to 50 C For Operation Use in interior spaces Altitudes of up to 3000 m above sea level Temperature 2 C to 40 C Max relat...

Страница 14: ...ation 0 2 kW h Environmental Conditions For Storage and Shipping Temperature 2 C to 50 C For Operation Use in interior spaces Altitudes of up to 3000 m above sea level Temperature 2 C to 40 C Max rela...

Страница 15: ...tion 0 3 kW h Environmental Conditions For Storage and Shipping Temperature 2 C to 50 C For Operation Use in interior spaces Altitudes of up to 3000 m above sea level Temperature 2 C to 40 C Max relat...

Страница 16: ...lectronic equipment EN 61010 1 3 1 Edition EN 61010 2 020 3rd Edition EN 61010 2 011 2nd Edition EN 61010 2 101 3rd Edition EN 61326 1 Class B EN ISO 14971 ISO 13485 North America FDA listed Product c...

Страница 17: ...nterference to radio communications However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to radio or television...

Страница 18: ...Spin Micro 17R 230 V 10 50 60 Hz 1 9 A 320 W 4 A 2 pole circuit breaker 16 A 75002463 Fisherbrand accuSpin Micro 17R 120 V 10 60 Hz 3 9 A 330 W 8 A 2 pole circuit breaker 15 A 75003246 Fisherbrand acc...

Страница 19: ...in Micro 17R R 134a 0 26 kg 21 bar 1430 0 37 t 75002463 Fisherbrand accuSpin Micro 17R R 134a 0 26 kg 21 bar 1430 0 37 t 75003247 Fisherbrand accuSpin Micro 21R R 134a 0 26 kg 21 bar 1430 0 37 t 75003...

Страница 20: ...rature 121 C Table 11 General Technical Data 24 x 1 5 2 0 mL Rotor Rotor Performance Data 17 Series Centrifuges 24 x 1 5 2 0 mL Rotor Voltage 230 V 120 V Maximum Speed 13300 13300 Maximum RCF Value 17...

Страница 21: ...Acceleration Braking Time 12 s 13 s 12 s 13 s Min Temperature at max Speed Ambient Temperature of 23 C 0 C 0 C Table 12 24 x 1 5 2 0 mL Rotor Performance Data Accessories Description Article No Rotor...

Страница 22: ...15 General Technical Data Dual Row 18 x 2 0 0 5 mL Rotor Rotor Performance Data 17 Series Centrifuges Dual Row 18 x 2 0 0 5 mL Rotor Voltage 230 V 120 V Maximum Speed 13300 13300 Maximum RCF Value 16...

Страница 23: ...f 23 C 0 C 0 C 21R Series Centrifuges Dual Row 18 x 2 0 0 5 mL Rotor Voltage 230 V 120 V Maximum Speed 14800 14800 Maximum RCF Value 20800 20800 Acceleration Braking Time 11 s 13 s 11 s 13 s Min Tempe...

Страница 24: ...General Technical Data 36 x 0 5 mL Rotor Rotor Performance Data 17 Series Centrifuges 36 x 0 5 mL Rotor Voltage 230 V 120 V Maximum Speed 13300 13300 Maximum RCF Value 15600 15600 Acceleration Brakin...

Страница 25: ...Temperature of 23 C 0 C 0 C 21R Series Centrifuges 36 x 0 5 mL Rotor Voltage 230 V 120 V Maximum Speed 14800 14800 Maximum RCF Value 19300 19300 Acceleration Braking Time 9 s 11 s 9 s 11 s Min Tempera...

Страница 26: ...erature 121 C Table 23 General Technical Data 10 x 5 mL Rotor Rotor Performance Data 17 Series Centrifuges 10 x 5 mL Rotor Voltage 230 V 120 V Maximum Speed 13300 13300 Maximum RCF Value 16414 16414 A...

Страница 27: ...ies Centrifuges 10 x 5 mL Rotor Voltage 230 V 120 V Maximum Speed 14800 14800 Maximum RCF Value 20326 20326 Acceleration Braking Time 12 s 13 s 12 s 13 s Min Temperature at max Speed Ambient Temperatu...

Страница 28: ...able 27 General Technical Data PCR 8 x 8 Rotor Rotor Performance Data 17 Series Centrifuges PCR 8 x 8 Rotor Voltage 230 V 120 V Maximum Speed 13300 13300 Maximum RCF Value 13800 13800 Acceleration Bra...

Страница 29: ...t Temperature of 23 C 0 C 0 C 21R Series Centrifuges PCR 8 x 8 Rotor Voltage 230 V 120 V Maximum Speed 14800 14800 Maximum RCF Value 17100 17100 Acceleration Braking Time 7 s 9 s 7 s 9 s Min Temperatu...

Страница 30: ...e 31 General Technical Data PCR 4 x 8 Rotor Rotor Performance Data 17 Series Centrifuges PCR 4 x 8 Rotor Voltage 230 V 120 V Maximum Speed 13300 13300 Maximum RCF Value 13100 13100 Acceleration Brakin...

Страница 31: ...aximum Speed 14800 14800 Maximum RCF Value 16200 16200 Acceleration Braking Time 11 s 13 s 11 s 13 s Min Temperature at max Speed Ambient Temperature of 23 C 0 C 0 C Table 32 PCR 4 x 8 Rotor Performan...

Страница 32: ...report it to the carrier and request a damage inspection Important Failure to request an inspection of damage within a few days after receipt of shipment absolves the carrier from any liability for d...

Страница 33: ...o compensate for an uneven surface Do not operate the centrifuge on carts or stand alone shelving that could move during operation or is inadequately sized for the centrifuge Be able to hold the weigh...

Страница 34: ...34 89 Transport and Set Up Figure 1 Safety Zone...

Страница 35: ...t the centrifuge the centrifuge door is closed CAUTION Hands can be squeezed by an open centrifuge door Always close the centrifuge door before you transport the centrifuge Before transporting a rotor...

Страница 36: ...h 2 Make sure that the power supply cable specification agrees with the safety standards of your country 3 Make sure that the voltage and frequency are the same as the figures on the nameplate 4 Make...

Страница 37: ...s You can pre temper the centrifugation chamber and the empty rotor before the centrifugation run starts using SNOWFLAKE button NOTICE This function is only available on refrigerated centrifuges PULSE...

Страница 38: ...h for balancing potential voltage discrepancies in the power grid After pressing the mains switch the display therefore may still flash up to 10 seconds 3 2 3 Audible Alarm Error Accompanying all erro...

Страница 39: ...the centrifuge lid Close the centrifuge lid by pressing down on it lightly in the middle or on both sides The locking mechanism engages to close the lid safely The lid should audibly click into place...

Страница 40: ...fuge spindle The two bars in the labeling on the upper side of the rotor a must be aligned with the retaining pin b of the centrifuge spindle a b 3 Let the rotor slide down slowly 4 Insert the Allen w...

Страница 41: ...be hot 55 C Be aware of this risk and proceed carefully when you change a rotor after a run or wait until the motor has cooled down Before installing a rotor Remove any dust foreign objects or residue...

Страница 42: ...1 Put the rotor lid on the rotor nut 2 Push the rotor lid down until you see and hear the lock click If the lid does not close at all or only by force make sure that the sealing rings are correctly i...

Страница 43: ...otor Lid Seals CAUTION Loose parts may damage the centrifuge When operating the rotor without the lid the seals are not fixed in their position and can damage the centrifuge Tube Caps Always close the...

Страница 44: ...rrosion 2 Inspect the centrifugation chamber and drive shaft for damage such as cracks scratches or traces of corrosion 3 Check the suitability of the rotor and other used accessories against the Chem...

Страница 45: ...um Loading Each rotor is designed to run with its maximum load at maximum speed The safety system of the centrifuge requires that the rotor is not overloaded The rotors are designed to work with subst...

Страница 46: ...maximum RCF value is related to the maximum radius of the tube opening Remember that this value is reduced depending on the tubes buckets and adapters used This can be accounted for in the calculatio...

Страница 47: ...3 5 2 Pre selecting Speed 1 Enter the desired value by pressing the MINUS PLUS buttons repeatedly until the desired value shows You can adjust the speed in 100 rpm increments 2 Press START to confirm...

Страница 48: ...entrifuge radius of the 24 x 1 5 2 0 mL Rotor 75003424 Refer to RCF Value Explained on page 46 for further information NOTICE Avoid if possible speed ranges close to natural resonances of the system R...

Страница 49: ...etermine the sample temperature in C Proceed as follows 1 Enter the desired value by pressing the MINUS PLUS buttons under the right display repeatedly until the desired value shows You increase or de...

Страница 50: ...the activated pre temp function 2 Enter the desired value by pressing the MINUS PLUS buttons under the right display repeatedly until the desired value shows You increase or decrease the temperature...

Страница 51: ...fuge Refer to Safety Zone on page 34 Persons and hazardous substances must be kept out of this safety zone while centrifuging Once the main switch has been turned on the rotor has been properly instal...

Страница 52: ...LSE spinning will start and continue until you release the key The centrifuge accelerates to and brakes at maximum power The pre selected value is ignored NOTICE The centrifuge accelerates to maximum...

Страница 53: ...p detail Large seal in outer groove of rotor body bottom detail small seal in groove of rotor collar 1 Lubricate all seals 2 Press the large seal in the outer groove of the rotor body top detail 3 Pre...

Страница 54: ...id Quick Test As a quick test it is possible to test the aerosol tightness of fixed angle rotors using the following process 1 Lubricate all seals lightly Always use the supplied grease when lubricati...

Страница 55: ...oroughly cleaning the rotors they must be inspected for damage wear and corrosion Make sure that sealing rings are still smooth not brittle nor otherwise damaged Some sealing rings are not autoclavabl...

Страница 56: ...re of 50 C If drying boxes are used the temperature must never exceed 50 C Higher temperatures could damage the material and shorten the lifetime of the parts 8 Inspect the rotor and accessories for s...

Страница 57: ...ifuge to the edge of the table 2 Pull the clip below the sucking grid and remove the filter unit completely by pulling it down 3 Remove the cumulated dust with a soft cloth 4 Inserting the filter agai...

Страница 58: ...agent Observe the safety precautions and handling instructions for the disinfection agents used 4 5 Decontamination You are responsible that the level of decontamination is achieved according to your...

Страница 59: ...ICE No chemical additives are permitted in the steam 4 7 Service Fisher Scientific recommends having the centrifuge and accessories serviced once a year by an authorized service technician The service...

Страница 60: ...on with a decontamination certificate WARNING Before shipping the centrifuge and accessories you must clean and if necessary disinfect or decontaminate the full system If you are not sure consult with...

Страница 61: ...com centrifuge For the countries of the European Union the disposal is regulated by the European Union s Waste Electrical Electronic Equipment WEEE Directive 2012 19 EC Mind the information on transpo...

Страница 62: ...The braking process lasts much longer than usual Proceed as follows 1 Wait until the rotor has stopped This can take several minutes Use the view port for visual confirmation 2 Pull out the power supp...

Страница 63: ...o not use any sharp tools aggressive liquids or fire to fasten the melting process If necessary use warm water to speed up the melting process 4 Remove the water from the centrifugation chamber 5 3 Tr...

Страница 64: ...ation to standstill Centrifuge lid was opened manually during the run Close centrifuge lid immediately Rotor stops with deceleration to standstill For further centrifugation you have to switch the ins...

Страница 65: ...respectively the filter unit of the cooled centrifuge After approx 60 minutes you can restart the instrument Observe the maximum permissible environmental temperature If the error message still shows...

Страница 66: ...technician Table 34 Troubleshooting 5 3 8 Information for the Customer Service If you need to contact customer service please provide the order no and the serial no of your device This information can...

Страница 67: ...ptoethanol S S U S M S S U S S U S S S S S S U S S S S S S A cetaldehyde S U U M U M U U U M M M S U S U A cetone M S U U S U M S S U U S U S U U U S S U U S M M S U U A cetonitrile S S U S M S S S U...

Страница 68: ...U U S M U U S S U S U S U A mmonium A cetate S S U S S S S S S S S S S U S S S S S S S S S S A mmonium C arbonate M S U S S S S S S S S S S S U U S S S S S S M S S S A mmonium H ydroxide 10 U U S U S...

Страница 69: ...S S S S S M S S S S S S M S S S A mmonium S ulfate U M S S S U S S S S S S S S S S S S S S S U S S U A myl A lcohol S M U S S M S M S S S S M U S M A niline S S U U S U S M S U U U U U U U S M U U S S...

Страница 70: ...d U S S M S S U S S S S S S S S S U S S S S S S S S S S C esium A cetate M S S S S S S S S S S S S S S S S M S S S C esium B romide M S S S S S S S S S S S S S S S S S S M S S S C esium C hloride M S...

Страница 71: ...romic A cid 50 U U U U U S U U S M U M S S U M S U M S C resol M ixture S S U S S U U U U U U U U U S S S S U S C yclohexane S S S S S S U S U S S U U U M S M U M M S U M M U S D eoxycholate S S S S S...

Страница 72: ...nate S S U S S S S S U S U S U S S S M S S S S S S D imethylsulfoxide S S U U S S S S U S S U S U U S S U U S S S S U U D ioxane M S U U S S M M S U U S U M U U M M M U S S S S U U F erric C hloride U...

Страница 73: ...S U U S M M S U U E thyl A lcohol 50 S S S S S S M S S S S S U S U S S S S S S S S M S M U E thyl A lcohol 95 S S S U S S M S S S S S U S U S S S M S S S U S M U E thylene D ichloride S U U S M U U S...

Страница 74: ...ofluoric A cid 10 U U U M U U U S S M U S S S S M S U U U H ydrofluoric A cid 50 U U U U U U U U U S U U U S S M M S U U U M H ydrochloric A cid conc U U U U U U M U M U U M U U U S U S U U U F ormald...

Страница 75: ...mo S ol S S S S S S S S S S S S S S S S S S S S S H exane S S S S S S S S U S U M U S S U S S M S U S S U S I sobutyl A lcohol M U S S U S U S S M S S S S S S S S I sopropyl A lcohol M M M U S S S S S...

Страница 76: ...S S S S S S S S S S S S S S S S S S S U S S S P otassium H ydroxide 5 U U S S S S M S S S S S U S S S S S S S M U M S U P otassium H ydroxide conc U U M U M M S S U M U U U S M M U U U U P otassium P...

Страница 77: ...n Viton K erosene S S S S S S U S M U S U M M S M M M S S U S S U S S odium C hloride 10 S S S S S S S S S S S S S S S S S S M S S odium C hloride sat d U S U S S S S S S S S S S S S S M S C arbon T e...

Страница 78: ...S S S S S S S S S S S S S S S S S S S M S S S M ercaptoacetic A cid U S U S M S S M S U U U U S U U S M S U S S S S M ethyl A lcohol S S S U S S M S S S S S U S U M S S S S S S S M S M U M ethylene C...

Страница 79: ...M S S M S M S S S L actic A cid 20 S S M S M S S S S S S S M S M S S S N B utyl A lcohol S S U S S M U S M S S S S M M S M S S N B utyl P hthalate S S U S S S S U U S U U U M U U S U S M M S U S N N D...

Страница 80: ...S S S S S S S S S S S S S odium D odecyl S ulfate S S S S S S S S S S S S S S S S S S S S S S S S S odium H ypochlorite 5 U U M S S M U S S M S S S M S S S S M S S S M U S M S S odium I odide M S S S...

Страница 81: ...kel S alts U S S S S S S S S S S S S S S S S S S M S S S O ils P etroleum S S S S U S S S S U U M S M U U S S S U S S S S O ils O ther S S S M S S S S U S S S S U S S S S S S M S O leic A cid S U S S...

Страница 82: ...U M U U U U U S U M U U S U U U M S P hosphoric A cid 10 U U M S S S U S S S S U S S S S S S S S S U M U S S P hosphoric A cid conc U U M M U S M S U U M M S S S M S M S U M U S P hysiologic M edia S...

Страница 83: ...S S S S S S S S S S S S S S S S S S S S S S ucrose A lkaline M S S S S S S S S S S S U S S S S S S S S M S S S S ulfosalicylic A cid U U S S S S S S S S U S S S S S S S S S U S S S N itric A cid 10 U...

Страница 84: ...S S S S M S M U U M M S ulfuric A cid 10 M U U S S U U S S M U S S S S S S S S S S U U U S S S ulfuric A cid 50 M U U U S U U S S M U U S U U M S S S S S U U U M S S ulfuric A cid conc M U U U U U M M...

Страница 85: ...d U U U S S U M S U S U U S M M S S U U S U U U M U T richloroethane S U M U U S U U U U U U U U U S U S S T richloroethylene U U U U S U U U U U U U U U S U U S T risodium P hosphate S M S S S S S S...

Страница 86: ...ne S S U S S S M U S U U U U U U M U M U U U S U M S U S Z inc C hloride U U S S S S U S S S S S S S S S S S S S S S S U S S S Z inc S ulfate U S S S S S S S S S S S S S S S S S S S S S S S S S C itri...

Страница 87: ...7 H Hematocrit Rotor 31 How to install a rotor 40 How to Operate a Rotor 40 How to remove a rotor 41 I Ice Formation 63 Information for the Customer Service 66 Intended Use 5 Items Supplied 32 L List...

Страница 88: ...Standards 16 Storage 37 T Technical Data 12 Technical Specifications 11 Transport and Set Up 32 Troubleshooting 62 Troubleshooting by Guide 63 U Unpacking 32...

Страница 89: ...isherbrand accuSpin Micro 21R Thermo Electron LED GmbH Zweigniederlassung Osterode Am Kalkberg 37520 Osterode am Harz Germany Distributed by Fisher Scientific Contact us today UK Responsible Person Li...

Отзывы: