accumet AE150 pH Benchtop Meter
10 of 28
www.fishersci.com
Meter Setup Functions
Use the Setup mode to customize meter preferences and defaults:
•
P1.0: Memory Clear (CLr)
•
P2.0: View electrode data (ELE)
•
P3.0: Select buffer set (bUF)
•
P4.0: Reset to factory default settings (rSt)
To Enter the Setup Mode:
1. Turn the meter off by pressing the
ON/OFF
key.
2. With meter powered off, press and hold the
MODE
key and press and release the
ON/OFF
key. The display will show the “SETUP” indicator.
a. If the display indicates that the meter is in the measurement mode, turn the
meter off and repeat step 2 until the “SETUP” icon is shown.
Note:
To exit the Setup mode at anytime without confirming changes, press the
CAL/MEAS
key until the measurement mode appears. You may have to press the
CAL/MEAS
key two
times before the meter reverts to the measurement mode.
P1.0 Clear Memory Program
Use this program to clear all memory values when you need to store a new series of values.
This avoids confusing the old values with the new ones. NO is the default setting.
1. With the meter powered off, press and hold the
MODE
key
and press and release the
ON/OFF
key. The display will
show the “SETUP” indicator.
2. The meter will go to program P1.0. “CLr” is shown in the
upper display and “P1.0” is shown in the lower display.
3. Press the
ENTER
key to access the program P1.0. The
lower display will show either “nO” or “YES” (depending on
last option selected or factory default of nO).
4. Press the
or
arrow key to toggle between the “nO” or
“YES” setting. Selecting “YES” will erase all memory
values.
5. Press the
ENTER
key to confirm the selection and return to
the setup mode.
6. To go to the next program, press the
or
arrow key or
to return to the measurement mode press the
CAL/MEAS
key.