7
For proper installation follow the instructions below:
Leveling the Unit
Adjustable legs at the front corners of the refrigerator
should be adjusted so that the refrigerator is firmly posi-
tioned on the floor and the front is raised just enough that
the door closes easily when opened halfway. Turn leveling
legs clockwise to raise and counterclockwise to lower, see
Fig. 1.
Explosion-Proof Refrigerator
Installation
IMPORTANT:
PLEASE READ CAREFULLY
This unit is for free standing installation only.
The appro-
priate materials and wiring methods must be used in
order to
comply with current NFPA No.
70 NEC.
for Class
I, Group C
and Group D Hazardous Locations and Local
Codes.
Determine the total amount of current presently being
used by other apparatus connected to the circuit that will
be used by this unit. It is critical that this added current
not exceed the rating of the fuse or circuit breaker in use.
Installation
Warning
If the unit is tilted in excess of 30
degrees, level the unit, then wait 12
hours before applying power.
Turn clockwise to raise, counter-
clockwise to lower.
Figure 1
Warning
BEFORE ANY ELECTRICAL
CONNECTIONS ARE MADE, BE
SURE THERE IS NO EXPLOSIVE
CONCENTRATIONS IN THE AREA.
Caution
Be sure the voltage supplied to the
refrigerator/freezer is equal to that
specified on the data plate.
Warning
For personal safety, this unit must be
properly grounded before use.
The National Electrical Code (NEC) must be referred to
for proper fusing and service conductor size and type.