background image

 
Centrific

â

 Centrifuge  

Model 228 

Cat. No.  04-978-228 (115V) 
 04-978-229 

(230V) 

 

The Model 228 Centrific Centrifuge is 

a portable, bench-top centrifuge 

designed for clinical, educational. and 

industrial applications. The Centrific 

Centrifuge features a 6-place 45

°

 

angle nylon rotor and nylon shields 

which will accommodate tubes ranging 

in size from 7 to 17mm in diameter 

and 100 to 133mm in length. This 

includes 7ml, 10ml, and 15ml 

Vacutainer* tubes. Additionally, Tube 

Adapters 4-978-23 are available to 

accommodate the smaller 10-13 mm 

diameter by 75 mm long tubes. 

 

The unit operates at a fixed speed of 

3300 rpm and develops an RCF force 

of 1380 x G. Sedimentation is 

controlled by centrifugation time. The 

motor is thermally protected and has 

permanently lubricated ball bearings.  

The cast iron base and vibration 

isolation mounting provide additional 

stability. 

 

The feet on which the centrifuge rests 

incorporate suction cups which adhere 

to smooth benchtops.  This serves to 

reduce the movement of the unit due 

to vibration.  In order to prevent 

damage to these feet, 

never pick the 

unit straight up off of the benchtop.  

Always slide the unit off of the edge 

of the work surface. 

 

The Centrific Centrifuge also features 

a 15 minute timer with a hold position 

to give the user the option of 

continuous operation. The time 

selector functions as the power switch, 

engaging the motor when the run time 

has been set and the lid is closed. 

 

The Model 228 has a lid latch to 

positively secure the lid closed. The 

latch is operated by a knob located at 

the front of the lid. The latch operates 

through a quarter turn. 

 

Warning: Always be sure that the lid is 

latched before operating the unit

 
 
*Vacutainer is a registered trademark of 
Becton-Dickinson Co. 

 

UNPACKING 

Each Fisher Centrific Centrifuge is 

shipped in a single carton. Carefully 

unpack the carton and check the 

contents against the packing list 

below. Should a discrepancy exist, 

notify your Fisher branch or 

representative. 

 

NOTE: If damage is observed, keep 

the shipment intact -- including the 

carton and all packing material -- and 

file 

claim with the final carrier.  

Usually the firm will send an 

investigator to ascertain liability. 

 

Qty Item 

 

          Part No. 

  1    Centrifuge Assembly     
          04-978-228  (115v)                50775 
     or  04-978-229  (230v)              50867 

    1    Rotor, Cat.No.  04-978-232      50779 

  6    Shield, Cat.No. 04-978-233      50780 

1

 

Thumbscrew  

          04-978-228  (115v)                38816 

   or  04-978-229  (230v)              50864 

  1    Instruction Manual 

            70906

 

 

INSTALLATION 

To install the Model 228 Centrific

 

Centrifuge, perform the following: 

 

1.  Place centrifuge base on a flat 

level surface. 

2.  Remove the six shields from their 

packing and carefully insert them 

into the six holes in the rotor. 

3.  Place rotor on drive shaft. Align flat 

on drive shaft with hole position "1" 

on rotor. 

4.  Secure rotor (with supplied thumb-

screw) onto drive shaft. 

CAUTION:

 

Apply finger-tight 

pressure only: do not use 
wrench or other tool. 

5. Be sure that the six shields are in 

place. Never run the unit without all 

shields in place

 
 

CONTROLS 

 

Lid-Power Interlock  

The lid interlock is designed to 

interrupt power to the motor whenever 

the lid is opened. This interlock is 

implemented by a leaf switch. The lid 

should not be used as an ON/OFF 

switch. 

 

Time Control 

The time control is the power switch 

for the motor. To start a run, the time 

control is simply turned to the desired 

centrifugation time. During the run, the 

timer rotates counterclockwise to 

 

indicate the time remaining. The user 

may also use the “hold” feature to 

make self-timed runs by placing the 

time control into its HOLD position. 

 

OPERATION 

Centrifuge operation involves simply 

loading the rotor, closing and latching 

the lid, and selecting a centrifugation 

time. The rotor must be properly 

loaded to achieve good sedimentation 

and vibration-free operation. Tubes of 

the same capacity and load must be 

placed 180 apart. Use water, if 

necessary, to balance the load if there 

is an odd number of tubes. Balance 

load to within 1.0 gram. 

 

NOTE: Never use shot, weights, or 

mercury in tubes or shields. 

 

Procedure 

To operate the centrifuge, perform the 

following: 

 

1.  Turn the time control to OFF. 

2.  Connect the unit to an appropriate 

power source. 

3.  Unlatch and open lid; place tubes 

into rotor in a balanced pattern. 

4.  Close and latch the lid. 

5.  Rotate the time control CW 

beyond, then CCW back to, 

desired setting. 

6.  When the rotor has stopped 

spinning, raise the lid and remove 

the tubes. 

 

Determining Relative 

Centrifuge Force

 

To determine the relative centrifugal 

force (RCF), RPM and tube tip radius 

must be known.  Different rotor loads, 

tubes, line voltage, and condition of 

the centrifuge will affect the RPM.  An 

access hole for a phototachometer is 

provided in the lid.  The tube tip radius 

is defined as the distance from the 

center of the shaft to the farthest end 

of the tube.  When installed in the 

rotor on the Model 228, this distance 

is fixed at 11.32cm with shields 

installed, but with no tube adapters or 

cushions. 

 

RCF is defined as the force acting on 

a sample being centrifuged relative to 

the force, due to gravity, that normally 

acts on it. Such force must take into 

account the tip radius of the installed 

tube and the RPM at which it's 

spinning. Assuming that the normal 

gravitational force is at sea level, the 

following formula expresses the 

strength of the gravity field to which a 

sample is subjected. 

 

 

 

 

 

 

 

 

 

 

 

INSTRUCTIONS

 

 

Отзывы: