Stemi 1000/2000/2000-C/2000-CS
FL S configuration with LUMATEC HBO 200
B 40-002 e 12/98
3-9
Remove microscope body with Stemi mount (16/4) and safety
ring (16/5) from stand N (16/3).
Attach jointed mount S (16/1) and FL S glare protector (16/2) to
the stand column.
Then attach microscope body to the stand column again.
Fig. 3-16
Place LUMATEC HBO 200 SUV-DC-P light source (17/2) in
position, connect foot switch (17/4) to the instrument rear,
connect the unit to the line via line cable (17/3) (also see the
LUMATEC HBO 200 operation manual and the safety notes
included!).
Connect the liquid light guide (17/1) to the front of the
LUMATEC HBO 200 light source (use liquid light guide of
1500 mm length with special connector for LUMATEC
HBO 200).
Screw temperature-resistant special excitation filter (18/3) into the
used focusing attachment (18/2 or 4).
Attach liquid light guide (18/6) to jointed mount S (18/5) via light
guide adapter 10/15 (18/1), attach focusing attachment (18/2)
and clamp it.
Fig. 3-17
Attach focusing attachment (18/4) directly to the jointed
mount S (18/5).
NOTE
For the installation of the light guide and for the filter change
please see the separate notes on page 3-10.
Depending on the used microscope version, attach FL-S barrier
filter mount (19/1) for the DV-version or FL S barrier filter mount
(19/2) for the SV-version together with the required barrier filter
slider (14/3) (for a detailed description please see Figs 3-8 and 3-
9).
Fig. 3-18
CAUTION
Only use the temperature-resistant special filter sets for
the LUMATEC HBO 200.
NOTE
For UV excitation, please remember to use the barrier filter
slider (protects your eyes from reflecting/scattered UV excitation
light)!
Switch on LUMATEC HBO 200 illuminator in accordance with the
instrument manual, align the luminous field diameter with the
required object field and optimize illuminance through focusing
and variations of the working distance.
Adjust glare protector (19/4) for glare-free microscopy.
CAUTION
Glare to the eyes and direct irradiation of the skin with
excitation light must be avoided. If excitation is not required,
close the shutter which is integrated in the LUMATEC HBO 200.
Fig. 3-19
Содержание Stemi 1000
Страница 1: ...1 2 Stemi 1000 2000 2000 C Stereo Microscopes Operating Manual ...
Страница 42: ...Stemi 1000 Stemi 2000 Stemi 2000 C 12 98 B 40 002 e 1 32 ...
Страница 64: ...Stemi 1000 Stemi 2000 Stemi 2000 C 12 98 B 40 002 e 4 4 ...
Страница 66: ...Stemi 1000 Stemi 2000 Stemi 2000 C 12 98 B 40 002 e A 2 ...
Страница 68: ...Stemi 1000 Stemi 2000 Stemi 2000 C 12 98 B 40 002 e A 4 ...