MMS
®
Inspection SPG
15
Repeatability Precision*
Based on Fischer factory calibration standards,
5 single readings per standard, 20 °C (68 °F)
specimen temperature and 20 … 25 °C
(68 … 77 °F) ambient temperature
10
0
… 100 µm: ≤ 1.5 µm
100 … 500 µm: ≤ 1.5 % of reading
0
.94,
…
1
3.94 mils: ≤ 0.06 mils
3.94 … 19.69 mils: ≤ 1.5 % of reading
*
The data for accuracy and repeatability may differ in harsh industrial environments
Influence
Curvarure
Probe unsuited for measurements on curved surfaces
Edge distance
No influence; for the measurement, the touchdown table of the measuring instrument must
rest completely on the surface
Scope of Supply
Gage; 2 batteries; USB cable type C to type A (1 m (39.4 inches)); calibration standard
set 605-308; guideline
Order Information
MMS Inspection SPG
Gage
Variant
Order no.
Interface
Memory capacity
Vibration
Start
606-034
USB
max. 10,000 measured
values in 1 batch
High
606-035
USB + WiFi
250,000 measured values
in 2500 batches
We recommend ordering the probe tip replacement kit at the same time
Spare parts/accessory for
MMS Inspection SPG
Product
Order no.
Description
Calibration
standard set
605-308
Glass base (606-306),
2 depth standards 300 µm/11.8 mils (605-305) and
100 µm/3.94 mils (605-307)
Probe tip
replacement kit 605-434
3 measuring tips, exchange tool 605-248