119
Calibration
16
Calibration
16.0.1 Purpose of the Calibration
A calibration is required in the following situations:
A new measuring application has been set up.
A product was imported.
If during a reference measurement
1
a deviation was encountered that is greater than
two to three times the standard deviation s.
The calibration sets the difference between nominal value and actual value to zero.
Actual value refers to the reading displayed by the instrument.
The nominal value is the value engraved on the calibration standard (engraved value).
Calibrating enables
secured measurements
:
If the nominal values of the used standards
can be traced to recognized material measures or reference materials, then the measure-
ment is
secured
through the calibration.
16.0.2 Performing the Calibration
You need calibration standards for a calibration, whose coating thickness and material
composition are known.
The standards determine the absolute value of the measurement result. For this reason
the standards have to be kept in a safe place.
Standards are only suitable, if they are stable. This means that no reactions occur
between the coatings, with the substrate or with the environment.
Before you start
The instrument must be at operating temperature before performing the calibration.
Perform the calibration at the same workplace as you will perform the measurements.
If you set up the instrument at another place, perform the calibration at the new place
again.
1. A reference measurement is a measurement on a reference sample with known properties.
Occasionally, a calibration standard may also be referred to as a calibration nor-
mal.
Содержание FISCHERSCOPE X-RAY 5000 Series
Страница 18: ...18 FISCHERSCOPE X RAY Switching the Instrument on and off Communication between Instrument and WinFTM...
Страница 22: ...22 FISCHERSCOPE X RAY Performing Manual Measurements Deleting Measurement Readings...
Страница 36: ...36 FISCHERSCOPE X RAY User Interface of the WinFTM Software The Spectrum Window...
Страница 40: ...40 FISCHERSCOPE X RAY WinFTM File Structure Product...
Страница 112: ...Measurement device monitoring for the Fischerscope X RAY Long term monitoring 112 FISCHERSCOPE X RAY...
Страница 118: ...118 FISCHERSCOPE X RAY Def MA Display the Measurement Mode...
Страница 124: ...124 FISCHERSCOPE X RAY Calibration...
Страница 142: ...142 FISCHERSCOPE X RAY Addendum Periodic Table of the Elements with X Ray Properties...
Страница 156: ...156 FISCHERSCOPE X RAY Addendum Assignment of the Electrical Connections...
Страница 182: ...182 WinFTM WinFTM SUPER For the Experienced X RAY User Comparative Overview with without WinFTM PDM...
Страница 183: ...WinFTM 183...