Entering Compensation Spectra
Def.MA
The window
Periodic system
appears.
8. In the window
Periodic system
click the element that the layer is composed of.
The element, the percentage and the density are displayed in the lower area of the
window
Definition of Elements New Layer.
In case of a complex layer (mixture of materials or alloy) enter more elements and their
percentages.
9. Click
Elem.+
to transfer the values to the display field.
The values of the entered layer appear in the large display field.
10. Repeat steps 8 to 9 for each element of the current layer.
11. Click
OK
to finish the value entry for the current layer and to save the data.
The window
Definition of Elements New Layer disappears.
The values of the layer
appear in the large display field of the window
Definition of measuring conditions... .
12. Repeat steps 4 to 11 for each layer of the coating system until all layers are entered.
Enter the base material (substrate material) like another layer with the following
options:
a
In the area
Thickness
check the option
The thickness of the layer is fixed (known)
.
b
Check the option
Massive (Sat.)
13. Click
OK
to close the window
Definition of measuring conditions...
.
What you can do next
You can enter compensation spectra, see Chap. 15.3 on Page 115.
You can create a new product based on the created Def.MA.
15.3 Entering Compensation Spectra
By entering compensation spectra you can increase the accuracy of the measurement.
Before you start
The WinFTM extension SUPER is activated.
The Def.MA is created, see Chap. 15.2 on Page 114.
Procedure
1. In the WinFTM main menu select
Calibrate > Def.MA modify...
.
The window
Def.MA modify
appears.
2. Select the Def.MA and click
OK
.
The window
Definition of measuring conditions...
appears.
3. In the window
Definition of measuring conditions...
click
Compens.Spectra
.
The window
Spectrum Background
appears.
4. You can select one or more of the following options:
Содержание FISCHERSCOPE X-RAY 5000 Series
Страница 18: ...18 FISCHERSCOPE X RAY Switching the Instrument on and off Communication between Instrument and WinFTM...
Страница 22: ...22 FISCHERSCOPE X RAY Performing Manual Measurements Deleting Measurement Readings...
Страница 36: ...36 FISCHERSCOPE X RAY User Interface of the WinFTM Software The Spectrum Window...
Страница 40: ...40 FISCHERSCOPE X RAY WinFTM File Structure Product...
Страница 112: ...Measurement device monitoring for the Fischerscope X RAY Long term monitoring 112 FISCHERSCOPE X RAY...
Страница 118: ...118 FISCHERSCOPE X RAY Def MA Display the Measurement Mode...
Страница 124: ...124 FISCHERSCOPE X RAY Calibration...
Страница 142: ...142 FISCHERSCOPE X RAY Addendum Periodic Table of the Elements with X Ray Properties...
Страница 156: ...156 FISCHERSCOPE X RAY Addendum Assignment of the Electrical Connections...
Страница 182: ...182 WinFTM WinFTM SUPER For the Experienced X RAY User Comparative Overview with without WinFTM PDM...
Страница 183: ...WinFTM 183...