6 | Technical data
FISCHER Mess- und Regeltechnik GmbH
22/36
BA_EN_DA03_SR_NM
Assembly
Wall mounting
Flanged assembly plate
Pipe mounting
Flanged assembly plate and attachment bracket
Panel mounting set
type 1
Panel installation fittings for units with a small measuring
cell (Ø75) and NG100 bayonet ring casing.
Panel mounting set
type 2
Front ring and support construction
6.4.1 Additional Attachments
6.4.1.1 Contact elements
Limit signal transmitters (contacts) and capacitive rotation angle transducers
with an output signal proportional to the angular position can be fitted into a
housing augmented by a corresponding bayonet ring connector.
A certain minimum pressure level is required to operate this kind of contact ele-
ment, which is why there is a lower limit for the mbar measuring ranges. This
limit depends on the model type and is stated in the section ‚General‘.
The measuring deviation increases by ±0.5% per contact when the contacts are
driven and switched.
For more information and the order key, please refer to the data sheet:
• for limit switch in data sheet KE
• for rotation angle converter in the data sheet KE09
6.4.1.2 Fluid charging
Under aggravated operating conditions, such as vibrations and extreme pres-
sure fluctuations, or in order to avoid condensation forming if used outdoors, the
casing can be filled with the following fluids depending in the type of contacts in-
stalled:
without contacts
Silicon oil
Low-action contacts
Silicon oil
Magnetic spring contacts
Silicon oil
Inductive contacts
no filling possible
Rotation angle transducer
no filling possible
6.4.1.3 Marker needle
A settable red marker can be attached to the scale to clearly show a certain
pressure (limit value).
6.4.1.4 Trailing needle
The railing needle is 'dragged' with the measured value indicator. As there is no
fixed connection between the two needles, one-off maximum values are stored.
The trailing needle can be reset using an adjusting dial in the window. Trailing
needles cannot be used in conjunction with contacts. A certain minimum pres-
sure level is required to move the drag indicator, which is why there is a lower
limit for the mbar measuring ranges. This limit depends on the model type and
is stated in the section ‚General‘.
6.4.1.5 Shut-off fitting
3-spindle valve block PN 420, DN 5, can be directly flanged
• Type DZ3600SV2700
• Material 1.4571
• Functions: Shut-off, pressure compensation