First Ten Angstroms FTA125 Скачать руководство пользователя страница 3

 

Start Auto

 means to start an automatic pumping operation, just as you would expect if you had 

checked 

Start on Run

.  You can also 

Run Pump Program now

 
 

 

 
 

Pump Tubing 

 
The Kloehn pump is shipped with 1/16” (1.6mm) push-on tubing barbs on both the Kloehn pump 
ports and the needle unions.  Normally this is used with Tygon tubing, as this tubing is 
inexpensive, flexible, and commonly available.  However, alternative tubing assemblies are 
available.  The fittings are ¼-28 threads.  Hamilton and Upchurch, to name two well-known 
suppliers, offer fittings and tubing using Teflon or PEEK.  FTA can also supply these.  If you use 
a stuffer tubing, such as Teflon or PEEK, you may wish to route the tubing along the front of the 
instrument frame and not make the cosmetic loop that the factory makes with Tygon tubing.  
Note the black Delrin cable and tubing guides are mounted into the frame slots with M5 screws 
and T-nuts and can easily be re-positioned.  Tygon tubing is sold by the major chemistry supply 
houses (Cole-Parmer, Fisher, VWR, etc.) and their catalogs describe its chemical compatibility. 
 
 

 

465 Dinwiddie Street 

 Portsmouth, Virginia 23704 

 1.757.393.1584  Fax: 1.757.393.3708 

http://www.firsttenangstroms.com 