Rear view of the Series A benchtop showing the power supply, fan, USB, Ethernet, reset button and serial number.
6.3
Environmental Conditions
Storage Conditions
The WaveShaper should be stored in appropriate dry, temperature controlled conditions to prevent damage to the equipment.
Exceeding the ranges defined in the table may result in permanent damage to the equipment.
Absolute Maximum Ratings
CONDITION / PARAMETER UNIT MIN MAX
NOTES
Storage Temperature
°
C
-20
+70
Storage Humidity
%
10
90
Non-condensing
Case operating temperature
°
C
15
55
Module Only
ESD
kV
16
Class N per HBM, meets IEC 61000-4-2 level 4
Exceeding the ranges defined here may result in permanent damage to the unit.
Normal Operating Conditions
CONDITION / PARAMETER
UNIT MIN MAX
NOTES
Operating Temperature
(module)
°
C
15
55
Temperature of case with airflow across top of unit. Min flow rate =
1m/sec.
Operating Temperature
(benchtop)
°
C
15
35
Air temperature measured at input to vents on base of unit
Humidity
%
10
90
Non-condensing
The device will operate within specification provided these operating conditions are met.
WaveShaper A Series User Manual
46