FineTek SIS2 Скачать руководство пользователя страница 1

 

 

 

 

 

SIS2 Intelligent 

Level Sensor 

Operation Manual 

 

 
FineTek Co., Ltd.

 

No.16, Tzuchiang St., Tucheng Industrial Park, New Taipei City 23678   

Tel: 886-2-22696789 

Fax: 886-2-22686682 

Website: http://www.fine-tek.com

 

E-mail:[email protected] 

08-SIS20XX-B1-EK,08/05/2021

Содержание SIS2

Страница 1: ...Sensor Operation Manual FineTek Co Ltd No 16 Tzuchiang St Tucheng Industrial Park New Taipei City 23678 Tel 886 2 22696789 Fax 886 2 22686682 Website http www fine tek com E mail info fine tek com 08...

Страница 2: ...orking Principle 8 7 1 Output status 8 8 Technical Performance 9 8 1 Specifications 9 9 External Dimensions 10 9 1 Dimensions 10 10 Installation 11 10 1 Installation environment 11 10 2 Installation p...

Страница 3: ...of the operation manual is updated based on the version upgrade which will be uploaded to the website for the user to access Please don t disassemble or repair the product on your own as this will mak...

Страница 4: ...ing to the actual condition The product or its parts are beyond the warranty period Fault or damage is caused by not following the instruction and use environment described on the operation manual The...

Страница 5: ...ine automation Co Ltd China No 451 Duhui Road Zhuanqiao Township Minhang District Shanghai City 201109 86 021 64907260 86 021 6490 7276 FineTek Pte Ltd Singapore Branch 37 Kaki Bukit Place Level 4 Sin...

Страница 6: ...stall due to the standard connector Protection rating IP67 The lightweight and compact design make it easy to carry and enable for quick installation even in narrow spaces or difficult situations Prob...

Страница 7: ...erly Limitations Not suitable for bulk solid or rough material such as quartz sand Not suitable for highly corrosive media strong acids and strong alkaline Please check the compatibility with the prod...

Страница 8: ...ifferent resonance frequency is created according to different medium Thus a switching signal will be triggered if the sensor is covered with material 7 1 Output status The following table shows the c...

Страница 9: ...Switch type Group 1 output NO Group 2 output NC Output load current Max 50 mA Output voltage drop Max 2 5 VDC Short circuit protection Short pulse Overload protection Yes Switch delay Standard 1 secon...

Страница 10: ...10 9 External Dimensions 9 1 Dimensions...

Страница 11: ...t the product is working properly via the field test function If interference occurs take appropriate measures shielding grounding For pipe installation the pipe size must be bigger than DN80 Install...

Страница 12: ...the sensor through the thread lock sleeve The seal must be correctly positioned and locked Close the gap between the sensor and the adapter sleeve end face An inappropriate seal may cause installatio...

Страница 13: ...ncluded in the order customized order if need When tightening the sealing ring is not exposed and must be compressed at least 1 4 wire diameter or more than 0 3mm to ensure the seal and the tightening...

Страница 14: ...20 incline will can avoid material bridges from forming It also won t be damaged by the inflowing material 2 For top mounted installation the rear mounted type with the extension of the auxiliary rod...

Страница 15: ...ake sure that the top cable is properly sealed such as installing and locking the cable gland Cables cannot be routed directly upwards please go around for a short distance and then route them upwards...

Страница 16: ...nections according to the wiring diagram below Insert the connector plug into the top socket of the sensor and tighten the nut For hexagon nuts please use a 14mm open spanner to tighten torque of 1 1...

Страница 17: ...2 30Vdc 2 Lean a magnet close to the sing for 3 seconds or more there will be switching output with corresponding LED light display 3 Remove the magnet from the sign the switching output and correspon...

Страница 18: ...xagonal head surface again for staying one second and remove the magnet The yellow and green lights will flash respectively for 1 second At this time the sensor will record this status for not touchin...

Страница 19: ...the sealing cone Take appropriate methods to ensure that no external matter enters the device Attention If the medium is changed you need to change the sensor model The sensor is packaged on a one tim...

Страница 20: ...ss than 50mA and make corrections Over temperature protection is active 40 85 C The alarm is automatically cleared if the temperature returns to normal The green light continues to flash once per seco...

Страница 21: ...ty of the detected medium Suitable Not suitable Item Scope 1 Flour 2 Starch 3 Cocoa powder 4 Coffee powder 5 Hazelnut powder about 40 C 6 Pepper ground 7 Mashed potato 8 Creamer powder 9 Salt 10 Caste...