Figure 2

Preparing the Preamplifiers

The stainless steel electrodes are screwed directly into the 

preamplifiers. This direct mounting technique eliminates artifact 

problems caused by movement of electrode lead wires and electrical 

interference. However, these electrodes can be removed from the 

preamplifiers and reconnected to them with lead wires having screw 

type (4-40 UNC) connectors thereby allowing for remote placement if 

so desired.
Ensure that the electrodes or any connector wires used are screwed 

snugly (finger tight only) into the preamplifier. Do not over-tighten.

Using the Myolab II to Monitor EMG Activity

1.   To avoid static damage to the preamplifiers, always ensure that

the patient is at the same electrical potential as the Myolab. Two 

good ways of doing this are:
a)  After properly picking up the preamp (ground first), touch the

patient with your free hand before placing the preamp on the 

site being examined.

b)  Have the patient touch the ground (center) electrode before

the active (outside) electrode comes in contact with the patient.

2.   After guarding the patient, place the preamplifier with its

electrodes on the skin directly over the muscle group to be 

monitored (Figure 3). Be certain that the three electrodes on 

each preamplifier are all in good contact with the skin. The 

use of a strip of hypoallergenic medical grade tape to hold the 

preamplifier in place is recommended.

Содержание Myolab II

Страница 1: ...Myolab II Prosthetist Manual...

Страница 2: ...2...

Страница 3: ...yoelectric prosthetic device senses EMG signals from intact muscles and translates these signals into usable arm movements The Myolab II uses this same advanced sensing technology to provide artifact...

Страница 4: ...plifiers are installed inside the socket of a prosthesis make sure that the socket provides adequate drainage for perspiration System Setup Changing Batteries Hold the Myolab II upside down with the b...

Страница 5: ...to Monitor EMG Activity 1 To avoid static damage to the preamplifiers always ensure that the patient is at the same electrical potential as the Myolab Two good ways of doing this are a After properly...

Страница 6: ...he EMG signal can occur if both preamplifiers are not in contact with the patient common ground during monitoring 4 Set the gain dials Gain A and Gain B to 0 5 Rotate the sound dials A and B to the le...

Страница 7: ...ired level using the appropriate sound dial Note The pitch of the sound not the volume will increase with an increase in EMG signal intensity In addition the sound of each channel differentiates the t...

Страница 8: ...he same function in the sound arm e g flexing the sound elbow and contracting the biceps simultaneously in both the remnant arm and Figure 5 the sound arm Figure 5 For both bilateral and unilateral am...

Страница 9: ...ic socket Identify all potential EMG control sites in this manner 6 When probing EMG control sites for the Utah Artificial Arm use the A channel of the Myolab II to monitor the muscle the A muscle mos...

Страница 10: ...ck feature of the Utah Artificial Arm instruct the patient to contract both the flexor and extensor muscles simultaneously in a quick but low strength co contraction followed by relaxation the motion...

Страница 11: ...control To measure the EMG signal magnitude accurately set the gain dial on 10 and the gain switch on X1 B The muscle EMG signals should be controllable without spasmodic jumping or quivering C Contra...

Страница 12: ...k all connections and then turn the power On again Check the gain switch setting Change the battery If the Myolab II continues to malfunction after you have taken the above actions call the service de...

Страница 13: ...voltage Bias Current 1 picoamp Filtering 1st order high pass at 10 Hz 1st order low pass at 10 kHz Output Impedance 1 Ohm Output Current 6 mA Output voltage 9 V supply voltage 7 5 V output T Cable to...

Страница 14: ...14...

Страница 15: ...15 Alternate leads for use with Myolab II...

Страница 16: ...2018 Motion Control Inc 1930001 Rev D 10 04 2018 F Motion Control Inc 115 N Wright Brothers Drive Salt Lake City UT 84116 801 326 3434 Fax 801 978 0848 Toll Free 888 696 2767 www UtahArm com...

Отзывы: