6
Technical Specifications
6.1
Frequency Bands and Measurement Range
Detection method
Time-domain sampling (4096 samples) with digital processing and filtering
Sensors
Tree axis TMR sensor; Hall sensors
Measurement interval
User-selectable; from 4 seconds to 6000 seconds in steps of 0.25 second
6.2
Bandwidth and Resolution
The measurement range of the device is subdivided into four bands: Three high sensitivity ranges
with variable frequency resolution cover the frequency range from DC to 1 kHz, 10 kHz, and 100 kHz,
respectively. An additional high field strength mode allows measurement up to 30 mT (RMS) from
DC to 1 kHz.
Band Name
Bandwidth
Resolution Bandwidth
Maximum Field Strength
LF
(high resolution)
DC
–
1 kHz
1 Hz
1000 μT (RMS)
MF
DC
–
10 kHz
10 Hz
1000 μT (RMS)
HF
(high bandwidth)
DC
–
100 kHz
100 Hz
1000 μT (RMS)
High-B
DC
–
1 kHz
1 Hz
30 mT (RMS)
6.3
Sensitivity
Band
Frequency Range
Spectral Noise Density
LF, MF, HF Band
16 Hz - 150 Hz
< 5 nT /
√Hz
(see noise floor spectra below)
LF, MF, HF Band
150 Hz
–
100 kHz
< 2 nT /
√Hz
(see noise floor spectra below)
High-B Band
DC
–
1 kHz
< 0.2 mT/
√Hz
(see noise floor spectra below)
Figure 7: Typical noise floor: LF band (DC
–