Fieldpiece SMAN4 Скачать руководство пользователя страница 2

11

16

12

17

13

18

14

19

15

20

Pulling a Vacuum

Follow all manufacturers’ evacuation procedures over those in 

this manual.

1  Connect your SMAN4 in-line with your vacuum pump and the 

system, then power on your SMAN4.

2  Draw a vacuum on the system. The SMAN4 will automatically 

sense the negative pressure and display the vacuum in microns.  

3  Press Alarm to activate the High (Hi) alarm or Low (Lo) alarm. No 

display of Lo or Hi means alarm is off.  Note:  The stopwatch will 

start when an alarm is activated. Pressing Alarm will reset the 

stopwatch.

4  When the rate of change in pressure is less than 1 micron per 10 

seconds, 

Stable

 will appear to the left of the micron reading. 

Additional Tips:

 

To achieve a deep vacuum as quickly as possible:

1  Use the shortest hoses with the largest diameter available.

2  Remove Schrader cores and core depressors. Core removal tools 

like the "MegaFlow Valve Core Removal Tool" can be purchased 

from Appion, Inc. to help with this process.

3  Inspect the rubber seals at both ends of your hoses for damage 

that may result in leakage.

4  Do not use hoses with low loss fittings when evacuating or 

pulling a vacuum on a system.

Set Vacuum Alarms

1  Hold ALARM for one second to enter Alarm Set mode.

2  Press ALARM to toggle between Alarm Hi and Lo.

3  Use UP or DOWN ARROW to select the value and press ENTER 

to lock in each digit. Once last digit in 

Alarm Lo

 is locked, your 

SMAN4 will automatically revert to normal testing mode.

SMAN4 to HVAC Guide® (HG3)

Your SMAN4 can wirelessly send meaurements to the HG3 System 

Analyzer (FW v 3.42 or higher) for deeper analysis and data logging. Your 

SMAN4 can send suction and liquid line pressures and temperatures, 

superheat/subcooling calculations, indoor wet bulb and outdoor drybulb 

temperatures with target superheat calculation (if SMAN4 is wirelessly 

receiving these measurements), and vacuum micron measurements 

(for datalogging) all in real-time. From the HG3, you can print reports 

on a PC to hand to customers and/or file for your own record-keeping.

The HG3 System Analyzer can receive wireless measurements from 

the SMAN4 to the following tests:  CheckMe!®, Superheat, Subcooling, 

Charging Jacket, and Data Logger.

1  Connect your SMAN4 to the system for measurement of suction 

and liquid line pressures and temperatures.

2  Sync wireless devices to SMAN4 to receive indoor wet bulb and 

outdoor dry bulb measurements. See Target Superheat section for 

wireless IDWB and ODDB instructions. 

3  Press circular button on SMAN4 to enter Superheat/Subcooling, T1 

T2 Direct, or Saturation temperature modes.

4  Hold the SYNC button until a beep is heard (>1 sec). Wireless 

signal bars icon in lower right corner of display will appear and 

begin searching.

5  Set HG3 to one of the following tests:  CheckMe!®, Superheat, 

Subcooling, Charging Jacket, or Data Logger.

6  With HG3 test, highlight a test line that corresponds to one of the 

measurements being sent from the SMAN4.

7  Hold SYNC button on HG3 until search pattern shows on right of 

test line. All receivable real-time measurements from the SMAN4 

will populate corresponding test lines on HG3.

8  Press OUTPUT on HG3 to view deeper analysis of the system.

Note:  The SMAN4 and HG3 must be within 10 feet (3m) for initial 

sync to occur. Once synced, the wireless range is 100 feet (30m).

Specifications

Wireless range:  

Up to 100 feet (30m)

Initial wireless sync range:  

Within 10 feet (3m)

Mini-USB port:

  For updating to newer versions of firmware

Display size: 

5 inches (diagonal)

Backlight: 

Blue (On for 1 minute unless turned off manually)

Battery:  

6 x AA (Battery life below based on alkaline type)

Battery life:

  Approx. 135 hours (without backlight and wireless)

Low battery indication:

 

  is displayed when the battery 

voltage drops below the operating level

Auto shut off:  

30 minutes of inactivity when APO is activated

Over range:  

"OL" or "-OL" is displayed

Operating environment:

 32°F to 122°F (0°C to 50°C) at <75%  RH

Storage temperature:

 -4°F to 140°F (-20°C to 60°C), 0 to 80% RH 

(with battery removed)

Weight: 

4.03 lbs (1.83 kg)

Pressure

Sensor type:  

Absolute pressure sensors

Connector type: 

 Standard 1/4" and 3/8" NPT male flare fitting

Range:

  29” HgV to 580psig (English), 74 cmHgV to 0 to 4000KPa 

(Metric), 4.000MPa (Metric), and 40.00bar (Metric)

Resolution:

 0.1 psi/inHg; 1 kPa/cmHg; 0.001MPa; 0.01bar

Accuracy:

  29” HgV to 0” HgV: ±0.2” HgV

74 cmHgV to 0 cmHgV: ±1 cmHgV

0 to 200 Psig: ±1 Psi; 0 to 1378 KPa: ±7 KPa; 0 to 1.378MPa: 

±0.007MPa; 0 to 13.78 bar ±0.07 bar

200 to 580 Psig: ±0.3%+1 Psi; 1378 to 4000KPa: ±0.3%+7 Kpa; 

1.378 to 4.000MPa: ±0.3%+0.007MPa; 13.78 to 40.00bar: 

±0.3%+0.07bar

Maximum overload pressure:

 800 psig

Units:

 Psig, kPa, MPa, bar, inHg, and cmHg

Microns for Vacuum

Connector type: 

 Standard 1/4" and 3/8" NPT male flare fitting

Range:

  50 to 9999 microns of mercury

Resolution:

 1 micron (50 to 2000 microns), 250 microns (2001 to 

5000 microns), 500 microns (5001 to 8000 microns), 1000 microns 

(8001 to 9999 microns)

Accuracy:

  ±10% or ±10 microns, whichever is greater (50 to 1000 

microns)

Maximum overload pressure:

 580 psig

Units:

 Microns of mercury

Temperature

Sensor type: 

 K-Type thermocouple

Range: 

-95°F to 999.9°F (-70°C to 537.0°C)

Resolution:

 0.1°F/°C

Accuracy: 

±(1.0°F) -95°F to 199.9°F; ±(0.5°C) -70°C to 93°C 

±(2.0°F) 200°F to 999.9°F; ±(1.0°C) 93°C to 537.0°C 

Note: All accuracies are after a field calibration.

Refrigerants

The P-T charts of the following 39 refrigerants come pre-

programmed into your SMAN4. In your SMAN4 the refrigerants are 

listed in order of most commonly used. Here, they are listed in numerical 

order for easy reference.

R11, R113, R114, R12, R123, R124, R125, R13, R134A, R22, R23, 

R401A(MP39), R401B, R402A, R402B, R404A, R406A, R407A, R407C, 

R408A, R409A, R410A, R414B (Hotshot), R416A, R417A, R420A, 

R421A, R421B, R422A, R422B(NU22B), R422C(Oneshot), R422D, R424A, 
R434A(RS-45), R438A(MO99), R500, R502, R503, R507A

Maintenance

Clean the exterior with a dry cloth.  Do not 

use liquid.

Battery Replacement

The battery must be replaced when the 

battery life indicator 

 is empty.  Turn your 

SMAN4 off and replace with 6 AA batteries.

Cleaning the Micron Sensor

Over time, the vacuum sensor of the SMAN4 

may become contaminated with dirt, oil, and 

other contaminents introduced from pulling 

vacuums. 

Never

 use an object such as a cotton swab to clean the sensor, 

you may cause damage to the sensor.

2  Open all knobs/valves, and cap all the ports except for the VAC 

port. 

3   Drop enough Isopropyl (rubbing) alcohol into the VAC port using 

an eye dropper or funnel so that it can flush out contaminents.

4   Cap VAC port and gently shake your SMAN4 upside down to clean 

sensor.

5   Turn right side up. Open a port to pour out the rubbing alcohol 

and open all ports to allow sensors to dry out. Drying usually 

takes about an hour.

Calibration

Temperature 

To calibrate your SMAN4 temperature 

thermocouples, adjust the pot on the front of 

the meter labeled T1 Cal or T2 Cal. The best way 

to calibrate is to match to a known temperature. 

Ice water is very close to 32°F and is readily 

available. Accuracies of one degree or better 

are obtained through field calibration.

1   Stabilize a large cup of ice water by stirring. Pure, distilled water 

will be the most accurate.

2   Press the circular button until your SMAN4 enters Direct 

Temperature (T1 T2) mode. 

3   Immerse the temp probe in ice water from T1 and adjust the 

T1 Cal pot with a flathead screwdriver and let it stabilize, keep 

stirring.

4   Repeat Step 3 for temp probe in T2.

Pressure Zeroing

To calibrate your SMAN4 pressure sensors to 

atmospheric pressure, ensure that your SMAN4 

is disconnected from any pressure source and at 

equilibrium with the ambient pressure. 

1  SMAN4 must be in Superheat/Subcooling, Saturation, or T1 T2 

Direct mode. 

2  Press the CAL Atmospheric Pressure button and your SMAN4 will 

set the zero point of pressure to the ambient pressure.

Advanced Pressure Calibration

Your SMAN4 has the ability to perform a linear 

adjustment of the pressure sensors based on 

refrigerant type, temperature, and pressure.

Note: If the P1 and P2 pressure readings are not at zero, perform a 

Pressure Zeroing adjustment before performing this advanced 

calibration.

1   Press circular button until your SMAN4 enters T1 T2 Direct mode.

2   Plug in a K-type thermocouple into T1. (A bead type 

thermocouple, like the ATB1, is recommended.)

3  Connect the SMAN4 to a refrigerant cylinder of a known, single 

refrigerant using an EPA approved service hose. Be sure to open 

both HIGH and LOW side valves on your manifold and cap the 

unused ports. (If caps are not available you can connect both 

ends of a refrigerant hose to the two unused caps. Note: Some 

refrigerant will remain in the hoses which will need to be 

recovered.)

4   Press the REFRIGERANT button to match the refrigerant of the 

cylinder you are using.

5   Attach bead-type thermocouple to the side of the cylinder using 

tape. It is recommended to attach in the middle of the cylinder. 

Important:  Let the temperature of the thermocouple stabilize 

to the refrigerant temperature for 1 to 2 minutes or until stable. 

Note: For best results, the refrigerant cylinder should have been 

stored in a stable ambient condition environment for at least 24 

hours.

6   Open the refrigerant cylinder. The pressure inside cylinder should 

now be displayed on both HIGH and LOW side pressure sensors.

7   Press the CAL Test Pressure button. 

Your SMAN4 refers to the built-in P-T charts 

to compare the temperature of the refrigerant 

to the vapor saturation temperature. If the 

pressures on your SMAN4 are within 3psi of the 

pressure corresponding to the vapor saturation 

temperature, the SMAN4 will adjust the pressure 

sensor linearity to match the P-T chart. This is 

a fine adjustment and requires you to strictly 

adhere to the procedure closely or you will 

receive an error message "Err".

Limited Warranty

This meter is warranted against defects in 

material or workmanship for one year from 

date of purchase. Fieldpiece will replace or 

repair the defective unit, at its option, subject 

to verification of the defect.

This warranty does not apply to defects 

resulting from abuse, neglect, accident, 

unauthorized repair, alteration, or unreasonable 

use of the instrument.

Any implied warranties arising from the sale of 

a Fieldpiece product, including but not limited 

to implied warranties of merchantability and 

fitness for a particular purpose, are limited to 

the above. Fieldpiece shall not be liable for loss 

of use of the instrument or other incidental or 

consequential damages, expenses, or economic 

loss, or for any claim of such damage, expenses, 

or economic loss. 

State laws vary. The above limitations or 

exclusions may not apply to you.

Obtaining Service

Call Fieldpiece Instruments for current costs 

on our fixed price warranty service. Send 

check or money order for the amount quoted. 

Send your digital manifold, freight prepaid, to 

Fieldpiece Instruments. Send proof of date and 

location of purchase for in-warranty service. 

The meter will be repaired or replaced, at the 

option of Fieldpiece, and returned via least cost 

transportation.

www.fieldpiece.com

© Fieldpiece Instruments, Inc  2012;  v14

!

  WARNINGS

DO NOT APPLY MORE THAN 800 PSI TO ANY PORT ON THE MANIFOLD.

FOLLOW ALL EQUIPMENT MANUFACTURER'S TESTING PROCEDURES 

ABOVE THOSE IN THIS MANUAL IN REGARDS TO PROPERLY 

SERVICING THEIR EQUIPMENT.

Отзывы: