SPECIFICATIONS

Display:

3¾ digit liquid crystal display (LCD) with a maxi-

mum reading of 3999.

Range

:

Manual ranging mode (SC44 & SC45), hold

button for 2 seconds to return to autoranging mode.

Overrange:

"OL" or “-OL” mark indication.

Auto power off:

10 minutes.

Operating environment:

32

°

F to 122

°

F at <70% R.H.

Storage temperature:

-4

°

F to 140

°

F, @ <80% R.H. with

batteries removed.

Accuracy:

Specifications good in ambient conditions of

74

°

F

±

8

°

F, <75% relative humidity.

Temperature coefficient:

0.1

×

(specified accuracy) per 

°

F.

(32

°

F to 66

°

F, 82

°

F to 122

°

F).

Altitude:

6561.7 feet (2000m).

Power:

Two AAA 1.5V batteries.

Battery life:

400 hours typical with alkaline (SC44),

300 hours typical with alkaline (SC45),
200 hours typical with alkaline (SC46).

Accessories:

One pair test leads, (2) 1.5V AAA batteries

(installed), carrying case and operating instructions. 

Safety:

Designed to meet IEC61010-1 (EN61010-1) and

IEC61010-2-032 (EN61010-2-032), CATIII 300V, class II
and pollution degree 2 indoor use comply with CE,
UL61010B-1, UL61010B-2-032.

Temperature (SC45 and SC46)

Range:

-30

°

F ~ 400

°

F

Resolution:

0.1

°

F

Accuracy:

±(1

°

F) 32

°

F ~ 120

°

F, 

±(1% + 1.5°F) -4°F ~400°F,
±(2% + 4°F) -30°F ~ -4°F,          

Sensor type:

K-type thermocouple

Capacitance (SC45 and SC46)

Range:

4

μ

F, 40

μ

F, 400

μ

F, 4mF

4.000mF = 4000

μ

F

Resolution:

1.0nF

Accuracy:

±

(3% rdg + 10 dgts) on 4

μ

F

±

(3% rdg + 5 dgts) on 40

μ

F to 400

μ

F

±

(5% rdg + 20 dgts) on 4mF (4000

μ

F)

Minimum input range:

>100nF

Safety: 

“dsc” symbol on LCD indicates that the

capacitor must be discharged before testing.

DC volts

Ranges:

400mV (except SC46), 4V, 40V, 400V,

600V

Resolution:

0.1mV (1mV for SC46)

Accuracy:

±

(0.5% rdg + 2 dgt)

Input impedance:

400mV > 100M

Ω

4V = 10M

Ω

40V~600V = 9.1M

Ω

Transient protection:

6kV for 10

μ

sec

AC volts (50Hz - 500Hz)

True RMS: 

SC46 only, 

Crest factor:

3

Ranges:

4V, 40V, 400V, 600V

Resolution:

1mV

Accuracy(SC44, SC45): 

±

(1.2% rdg + 5 dgts), 

±

(1.5% rdg + 5 dgts) on 600V range

Accuracy(SC46): 

±

(1.2% rdg + 8 dgts), 

±

(1.5% rdg + 8 dgts) on 600V range

Input impedance:

4V = 10M

Ω

40V~600V = 9.1M

Ω

Resistance

Ranges:

400

Ω,

4k

Ω

, 40k

Ω

, 400k

Ω

, 4M

Ω

, 40M

Ω

SC45 and SC46 only have a  400

Ω

range.

Resolution:

0.1

Ω 

(1

Ω

for SC45, SC46)

Accuracy:

±

(1.0% rdg + 4 dgts) on 400

Ω

to 400k

Ω

ranges

±

(1.5% rdg + 4 dgts) on 4M

Ω

range

±

(3.0% rdg + 5 dgts)  on 40M

Ω

range

Open circuit volts:

-0.45VDC typical, (-1.2VDC on

400

Ω

range)

AC Current (50Hz - 60Hz)

True RMS: 

SC46 only,  

Crest factor:

3

Ranges:

40A, 400A, 

Resolution:

0.01A

Accuracy(SC44, SC45): 

±

(2.0% rdg + 6 dgts)

Accuracy(SC46): 

±

(2.0% rdg + 10 dgts)

Continuity

Audible indication:

Less than 25

Ω

Response time:

500ms

Diode Test

Test current:

1.2mA (appoximate)

Accuracy:

±

(3.0% rdg + 3 dgts) 

Resolution:

10mV

Audible indication:

Less than 0.25V

Open circuit volts:

3.0VDC typical

Voltage

Amps AC Through Clamp

Resistance

Capacitance

CAT.III

CLAMP

400A

300V

NCV

Disconnect test

leads from voltage
before plugging in
thermocouple!

Insure the

temperature being
measured is stable.
Maintain good
contact between the
thermocouple and
what's being
measured. 

WARNINGS

DISCONNECT AND UNPLUG TEST LEADS

before opening case.

TEST NCV FUNCTION ON KNOWN LIVE

WIRE before using.

DO NOT APPLY VOLTAGE greater than

30VAC or 60VDC to the thermocouple or the
jacks when the rotary dial is on OFF. 

REMOVE  THE  THERMOCOUPLE when

taking voltage measurements.

DISCONNECT THE  TEST LEADS when

taking temperature measurements. 

DO NOT APPLY VOLTAGE TO THE JACKS

when the rotary dial is on microamps. Even low
voltages can cause a current overload and blow
the fuse. Replace blown fuse to regain function.

!

!

CAT

.III

CLAM

P

400

A

300

V

NCV

Temperature 

VAC/DC

600VDC or AC RMS

AAC

400 AAC

Resistance

500VDC or AC RMS

Continuity

500VDC or AC RMS

Diode

500VDC or AC RMS

Capacitance 
(SC45 and SC46)

500VDC or AC RMS

Temperature 
(SC45 and SC46)

60VDC or 30VAC RMS

Overload Protection (SC44, SC45 and SC66)