Using the TruFirm 

The  TruFirm  Model  measures  the  depth  the  plunger  de-
presses a surface when it is released from a set height. The 
value of this depth is displayed on an LCD readout. 

The  unit  will  also  display  the  average  of  a  series  of  meas-
urements and the number of measurements included in the 
average on its LCD. 

Activating/Deactivating the Display: 

The  unit  is  activated  by  briefly  pressing  the  Button.  The 
LCD  will  display  the  percent  battery  life  for  5  seconds  and 
then show zeros when it is ready to take a measurement. 

The TruFirm will power off after 5 minutes of inactivity. 

Taking a reading: 

1.  Place the unit on the surface being measured. If the sur-

face  is  sloped,  orient  the  Base  so  that  it  is  pointing 
downhill. 

2.  Step lightly on the Foot Support. This will ensure the unit 

does not tip over after the reading is taken. 

3.  If the Display is blank, press the Button briefly and wait 

for the unit to turn on. 

4.  Lift  up  the  Plunger  all  the  way.  The  display  will  now 

show  the  number  of  measurements  that  have  been  in-
cluded in the average (or zero for the first reading). 

5.  Release the Plunger so that it drops smoothly. 

6.  On  the  first  measurement,  the  value  of  the  current 

measurement  is  displayed.    For  subsequent  measure-
ments,  the  LCD  will  display  the  current  reading  for  2 
seconds and the average after 2 seconds. 

To reset the average, press the Button briefly while the av-
erage  value  is  being  displayed.  If  the  average  is  not  reset, 
the  next  reading  will  be  included  in  the  average  as  well.  
The  average  will  also  be  reset  if  the  meter  is  powered  off. 

Содержание TruFirm

Страница 1: ...TruFirm Turf Firmness Meter PRODUCT MANUAL...

Страница 2: ...eld Scout TruFirm Turf Firmness Meter Please read this manual thoroughly before using your instrument Table of Contents General Overview 3 Specifications 3 Component Part Names 4 Setup 5 Using the Tru...

Страница 3: ...process the signal and send a measurement to an LCD display The measurement may also be streamed to a handheld device via a wireless communications protocol using a Bluetooth accessory items 6490S 649...

Страница 4: ...4 Component Part Names Actuator Arm Button Collar Plunger Shaft Base Foot Support...

Страница 5: ...d They are installed inside the plastic housing Remove the four screws holding the lid in place to access the battery holder Note If the TruFirm will not be used for an extended period of time over on...

Страница 6: ...orient the Base so that it is pointing downhill 2 Step lightly on the Foot Support This will ensure the unit does not tip over after the reading is taken 3 If the Display is blank press the Button br...

Страница 7: ...7 Release smoothly to avoid affecting drop speed Lift plunger straight up to top of travel Use foot support to stabilize the unit...

Страница 8: ...o store firmness measurements on multiple courses multiple holes multiple surfaces green fairway etc and at multiple times sessions It will also allow you to define additional surfac es from those pre...

Страница 9: ...end each measurement A separate FieldScout Mobile manual is available for the app Pairing the TruFirm Bluetooth Smart module to a smart device The TruFirm meter and the smart device must be paired to...

Страница 10: ...the plunger raised 0 06 This can be accomplished by placing a flat object about the thick ness of 15 sheets of paper under plunger The collar should just touch the actuator arm in its up most position...

Страница 11: ...11...

Отзывы: