18. Error Codes
Error Code No.
Description
Err 1
Time value error
Err 2
Date value error
Err 3
Exceed manual zero
Err 4
Offset out of range / unstable during power on (5 minutes for OIML and NTEP mode)
Err 5
No load cell detected
Err 6
Tare operation error
Err 7
Logic error. HI limit set is lower than LO limit (and HI is not = 0)
Err 8
Logic error. LO limit is higher than HI limit (and HI is not = 0)
Err 10
(F25 Mode 3) Container weight is higher than tArE H
Err 11
(F25 Mode 3) Container weight is lower than tArE L
Err 12
(F25 Mode 3 and Mode 4) Actual weight is higher than SP3
Err 13
Exceed maximum power on (5 minutes for OIML and NTEP mode)
Err 19
Capacity or division setting error (Division set is higher than 10000d)
Err 22
Manual Zero and Tare stability error
Err 23
Capacity setting error, Capacity 1 > Capacity 2
Err 24
Division setting error, e1 > e2
Err 25
Span gain is too low
Err 26
Not able to obtain stable status for longer than 10 sec
Err 27
Theoretical calculated value per e of Cal 2 varies more than 1% as of Cal 1.
Properly a load cell problem.
Mass value of Cal 2 is less than 150% of Cal 1.
Err 28
Maximum accumulation limit is exceeded.
Err 29
Standard deviation data exceed memory size (300 transactions).
--oL--
Overload (Gross weight is more than Max plus 9d)
HALT
Major system error detected. Power off instrument and remove power adaptor
immediately. Then check load cell connection and system power status.
UndEr
Negative Weight values exceeds display range
Reboot
Important parameters have been changed. Power off and then power on instrument
again to reboot.
------
Negative Tare value exceeds display range
19. Daily Care & Maintenance
a.
Clean the instrument with a soft, damp cloth. If necessary, use a mild detergent in water.
b.
Do not use any harsh, abrasive material, acetone, volatile solvent, thinner or alcohol for cleaning.
c.
Verify the accuracy of this instrument periodically. Re-calibrate if necessary. In some countries, calibration
requires authorized/qualified agent. Contact your dealer for more information.
d.
Store this instrument in a dry and clean place,
e.
Recharge battery before and every 4 months during long time storage.