FIBRO PGX+ PG1010 Скачать руководство пользователя страница 4

|   4

1.9 

Syringe Designs

The PG Dosing Unit has been designed with disposable low cost syringes in 
mind to avoid cleaning and contamination issues. These disposable syringes 
are equipped with a rubber plunger, which may not give exactly the same 
volume for each new pump cycle. Other very accurate syringes will fit into 
the PG Dosing Unit such as P/N860325, which contains five syringes with 
PTFE plunger and PTFE tubing with a steel cannulus fitted to the free end.
Use of a PTFE or glass plunger is highly recommended for measurement of 
wetting hysteresis (advancing/receding contact angles) where the pump 
flow is reversed.

  Please note the droplet should reach the surface and then it should  
  advance sufficiently across the surface without reaching the edge of the  
 image.
c) Press the REV button one or more times to reverse the pump flow. Please  
  note the droplet should now be receding until the liquid has disappeared  
  well inside the dispensing tip.
d) Press FWD again until the liquid appears at the dispensing tip and is ready  
  for a new pump cycle.
e) Press REV to finish the programming cycle.

1.6 

Liquid purging

If the PG Dosing Unit has been left standing for a short time, some of the 
droplet volume might be missing at the dispensing tip. Switch the mode 
selector to position ‘F’ (Fill) and press momentarily on the FWD button until 
the liquid reappears at the dispensing tip.

1.7 

LED Indicator messages

Off  

No power connected

Green  

Ready to operate

Green Blink   Pumping
Orange  

No program entered for this mode

Orange Blink   Programming in progress
Red  

Syringe is empty

1.8 

Functions

Mode 

 Button  

Function

F(Fill)  

(FWD)  

Purges a small volume when pressed  

  

momentarily

 

(REV)  

Reduces the droplet volume

 

FWD  

Pumps out liquid until released

 

REV  

Backs the syringe to its end position if kept  

  

pressed

 

 

-Abort operation by pressing FWD or REV

2, 4, 8  

FWD  

Pumps out a droplet of the preset size 

 

 

(e.g. 4 μL)

S1,S2,S3

   FWD  

Pumps out a droplet of programmed size

 

Programming

 

REV+FWD  

Enters programming mode

 

FWD*  

Enters the requested volume

 

REV  

Exits programming mode

C1,C2,C3

   FWD  

Performs a full pump cycle as programmed

 

Programming

 

REV+FWD  

Enters programming mode

 

FWD*  

Enters the requested advancing volume

 

REV*  

Enters the requested receding volume

 

FWD*  

Enters the requested ’top-up’ volume

 

REV  

Exits programming mode

* Can be pressed repeatedly to increase the volume setting

www.fibro-system.com