5.
Instrument maintenance and trouble shooting
5.1
Cleaning of fiber connectors
The optical output interface of this series of fiber optic comprehensive tester is a universal
replaceable interface. In case the instrument fails to test normally, or the test result is not
accurate, consider cleaning the interface connector first.
When cleaning, be sure to do it when the OTDR function and visual red light fault location
function are turned off. Unscrew the output port and wipe the end face with a woven lint free
clean wipe moistened with alcohol, then follow with a dry clean woven lint free wipe. Or utilize
ferrule rotating end face cleaner for a strictly dry clean procedure.
After cleaning, cover the port with dust cap until next use of the instrument and always keep it
clean.
5.2
Instrument screen cleaning
The display of this series of OTDR is a 4.3-inch TFT color LCD with touch screen. Do not touch the
LCD screen with sharp objects when using, if so the LCD screen may be damaged. When
cleaning, wipe the LCD screen with a soft woven cloth. Do not use organic solvent to wipe the LCD
screen as it may cause damage.
5.3
Calibration
It is recommended to calibrate the OTDR every 12 months. For calibration, please contact Fiber
Instrument Sales Inc. Phone: 800-500-0347 or 315-736-2206.