Handling and Storage

Storage: Store the microTargeting™ Electrode Lead Cable at normal temperatures between 0°C 
(32°F) and 40°C (104°F). Do not expose to relative humidity of more than 95%.

Specifications

C0230 – 3m Patient Lead

C0231 – 1.5m Patient Lead

C0232 – 1.5m Patient Lead (Special Use)

Note: C0232 is cleared for use in the US only. 

Length: 3 meters (C0230) / 1.5 meters (C0231, 
C0232)

For Single Use Only, do not re-sterilize

Pre-sterile, sterilization method – EtO

Minimum bend radius: 2cm

Configuration:  Dual, shielded Coaxial, 2.5mm OD

Interface Connector: proprietary, 6-pin

Electrode Connectors:

•  Color coded, red - macro, blk - micro
•  Gold plated (C0230, C0231)
•  Alligator Clips (C0232)
•  0.9mm ID x 3.5mm dp
•  Insertion/Extraction force: <500g

Reference Connector: Green, mini-alligator clip
Color Coded Flags: white, blue, green, red, yellow

Instructions for Use

The electrode lead cables are designed for use with the microTargeting™ Guideline 4000™ 5.0. For 
instructions on operating the Guideline system, refer to the Guideline 4000™ 5.0 Directions for Use, 
L011-85. The procedure below commences once the microelectrodes have been inserted.

Connecting an electrode to the Guideline 5 System is a two-step process:  connect the 
Patient Lead from the electrode to the UE Interface and ‘map’ that connection in the Guideline 
application.  While these steps may be performed in any order, to avoid the risk of making an 
error, they should both be completed for every electrode before moving on to the next one.

Connecting the Patient Lead: Maintaining sterile protocols, peel open the Patient Lead pouch.   
Connect the lead contacts to the electrode as shown:  black to black for the microelectrode 
contact, red to gray/red for the macro-electrode contact and connect the patient reference 
alligator clip to the insertion tube or any other suitable patient reference source. Being careful 
not to touch the non-sterile UE Interface (or handing off to a non-sterile assistant), plug the 
opposite end of the lead into the proper channel of the Interface.  It will be the one with the 
green status LED.

* Setup may vary for Special Use cables depending on electrode. Place the patient leads on the appropriate section of the 
electrode model being used. 

C0230/C0231 Setup

C0232 Setup*

Содержание microTargeting C0231

Страница 1: ...s C0230 C0231 C0232 FHC Inc 1201 Main Street Bowdoin ME 04287 USA Fax 1 207 666 8292 FHC Europe TERMOBIT PROD srl 42A Barbu Vacarescu Str 3rd Fl Bucharest 020281Sector 2 Romania 24 hour technical service 1 800 326 2905 US Can 1 207 666 8190 FHC Latin America Calle 6 Sur Cra 43 A 200 Edificio LUGO Oficina 1406 Medellín Colombia www fh co com ...

Страница 2: ...e European Community European Conformity This device fully complies with MDD Directive 93 42 EEC and legal responsibilities as a manufacturer are with FHC Inc 1201 Main Street Bowdoin ME 04287 USA Indicates a medical device that has been sterilized using ethylene oxide Indicates the temperature limits to which the medical device can be safely exposed Indicates the range of humidity to which the me...

Страница 3: ...ine 4000 5 0 Directions for Use L011 85 The procedure below commences once the microelectrodes have been inserted Connecting an electrode to the Guideline 5 System is a two step process connect the Patient Lead from the electrode to the UE Interface and map that connection in the Guideline application While these steps may be performed in any order to avoid the risk of making an error they should ...

Страница 4: ...an be matched to the colors used within the Guideline Application to represent that channel Associating a color in this way with each channel track electrode can be very helpful in keeping straight which channel is which throughout the procedure When the MER procedure is complete remove the lead cable from the electrode and Guideline 5 UE Interface Dispose of according to hospital protocol Recomme...

Отзывы: