Page 3 of 4
L011-85-05
Rev B0
2019-06-26
Handling and Storage
Storage: Store the microTargeting™ Electrode Lead Cable at normal temperatures between 0°C
(32°F) and 40°C (104°F). Do not expose to relative humidity of more than 95%.
Specifications
C0230 – 3m Patient Lead
C0231 – 1.5m Patient Lead
C0232 – 1.5m Patient Lead (Special Use)
Note: C0232 is cleared for use in the US only.
Length: 3 meters (C0230) / 1.5 meters (C0231,
C0232)
For Single Use Only, do not re-sterilize
Pre-sterile, sterilization method – EtO
Minimum bend radius: 2cm
Configuration: Dual, shielded Coaxial, 2.5mm OD
Interface Connector: proprietary, 6-pin
Electrode Connectors:
• Color coded, red - macro, blk - micro
• Gold plated (C0230, C0231)
• Alligator Clips (C0232)
• 0.9mm ID x 3.5mm dp
• Insertion/Extraction force: <500g
Reference Connector: Green, mini-alligator clip
Color Coded Flags: white, blue, green, red, yellow
Instructions for Use
The electrode lead cables are designed for use with the microTargeting™ Guideline 4000™ 5.0. For
instructions on operating the Guideline system, refer to the Guideline 4000™ 5.0 Directions for Use,
L011-85. The procedure below commences once the microelectrodes have been inserted.
Connecting an electrode to the Guideline 5 System is a two-step process: connect the
Patient Lead from the electrode to the UE Interface and ‘map’ that connection in the Guideline
application. While these steps may be performed in any order, to avoid the risk of making an
error, they should both be completed for every electrode before moving on to the next one.
Connecting the Patient Lead: Maintaining sterile protocols, peel open the Patient Lead pouch.
Connect the lead contacts to the electrode as shown: black to black for the microelectrode
contact, red to gray/red for the macro-electrode contact and connect the patient reference
alligator clip to the insertion tube or any other suitable patient reference source. Being careful
not to touch the non-sterile UE Interface (or handing off to a non-sterile assistant), plug the
opposite end of the lead into the proper channel of the Interface. It will be the one with the
green status LED.
* Setup may vary for Special Use cables depending on electrode. Place the patient leads on the appropriate section of the
electrode model being used.
C0230/C0231 Setup
C0232 Setup*