TYLOS

1 – Installing the patient making sco-

pic control possible (in profi le)

INDICATIONS 

The TYLOS™ nail is used in cases of combined talocrural and subtalar arthrodeses, 
as well as in certain forms of distal pseudarthrosis of the tibia when the ankle joint 
cannot be preserved.

Length: 160, 200, 250 mm

Side locking with 
a cortical screw

Posterior locking with a wedge-
type trans-calcaneal screw

The TYLOS™ ankle arthrodesis nail from the Be POD™ range has 
a unique posterior locking system, thanks to implantation of a 
trans-calcaneal screw of the threaded wedge type. This gua-
rantees considerable stability and thus reduces the risk of 
disassembly of the system or migration of the screw.

INSTALLING THE PATIENT

The patient is placed in dorsal decubitus. 

The leg to be operated is moved out of line and 
raised in relation to the controlateral leg so as to 
allow profi le controls with the X-ray image inten-
sifi er 

(Fig. 1)

.

• 

Cannulated nail, available in :

 

-

 3 lengths : 160, 200, 250 mm 

 

-

 3 diameters : 10, 11, 12 mm

 Material : Stainless steel

• 

Multi-axial stabilisation

 

-

  1 single wedge-type trans-calcaneal screw guarantees 

excellent mechanical stability – 

POSTERIOR

 

FIXATION

 

-

 3 to 4 laterally-attached cortical screws – 

LATERAL

 

FIXATION

 Wide range of trans-calcaneal and cortical screw sizes

 Protective plug

 Simple and precise instrumentation

 Drill guides designed for easy control 

   using X-ray image intensifi er

Содержание be POD TYLOS Arthrodese-Nail

Страница 1: ...FOOT TYLOS A N K L E ARTHRODESIS NAIL...

Страница 2: ...e stability and thus reduces the risk of disassembly of the system or migration of the screw INSTALLING THE PATIENT The patient is placed in dorsal decubitus The leg to be operated is moved out of lin...

Страница 3: ...will only be introduced into a hindfoot with a normal axis SUBTALAR APPROACH This is indispensable It is perfectly do able using a mini invasive approach via the lateral orifice made in the projectio...

Страница 4: ...mbled on the main fixation nail holder ref 253246 to which it is joined by the M6 connecting screw ref 253251 Depending on whether it is a right R or left L foot the notch R or L in the holder is plac...

Страница 5: ...the graduation desired and makes precise drilling possible as defined beforehand If the position of the nail does not make it possible to guarantee a minimum distance of 10 mm between the nail and th...

Страница 6: ...has remained in place and which allows cortical drilling Fig 16 Direct reading of the graduations on the drill bit blocked on the second cor tical determine the length of the screw required If you est...

Страница 7: ...ed in backwards making it initially possible to connect it to the screw and then when it reaches its blocking point to unscrew it in a second stage Fig 20 For the nail once you have returned to the pl...

Страница 8: ...alcaneal screw 254 590 lg 30 35 254 591 lg 30 45 254 592 lg 30 55 254 593 lg 40 35 254 594 lg 40 45 254 595 lg 40 55 FH ORTHOPEDICS References 253 239 5 lg 25 253 240 5 lg 30 253 241 5 lg 35 253 242 5...

Отзывы: