The IR only test does not use the UV bulb in the TalentumTT
2
and is the recommended test for IR2 and IR3 Talentum Flame
Detectors.
After selecting
IR Only Test
from the menu the TalentumTT
2
will prompt for confirmation:
Testing an IR2 or IR3 Talentum
The IR bulbs are turned on full power for several seconds to warm
them up to operating temperature.
Operating
29s
Cancel
+
Warming up
5s
Cancel
+
IR only Test
Confirm:
Aim torch at Talentum
within 2-8m. Do not
aim at eyes!
Press the
Trigger
button to start the test, or
Back
to cancel.
When the test starts the screen will change to show:
The number on the bottom row will count down in seconds. It is
not expected that the Talentum Flame Detector will react to this
stage of the test.
You should use this time to ensure you have the correct
alignment and distance between the TalentumTT
2
and the Flame
Detector under test as per ‘Setting up a Talentum Flame Detector
test’
on
page 14
.
Once warming is complete, the test moves onto the main
operation phase, and the screen will update to display: